BLADE Scale

4.  Once the weight is accepted, the display will show 

“PASS” then return to normal operation.

IX. Troubleshooting

• 

 

Problem:

 Scale will not turn on

• 

 

Solution: 

Check the batteries

• 

 

Problem: 

Displayed weight fluctuates randomly

• 

 

Solution: 

Make sure your workspace is stable and    

 free of any air currents or vibration

• 

 

Problem:

 Displayed weight is inaccurate

• 

 

Solution:

 Perform a calibration adjustment

X.  Error Codes

1.  OUTZ

 - Zero range has shifted. Perform calibration 

adjustment. If the problem persists, the load cell may 

be permanently damaged due to overloading.

2.  O-Ld

 - Maximum capacity exceeded. Remove the 

extra weight to avoid damaging the load cell.

3.  L 

- Batteries are low. Replace the batteries.

Summary of Contents for Blade-100

Page 1: ...BLADE 100 100g x 0 01g BLADE 650 650g x 0 1g BLADE 1KG 1000g x0 1g User Manual BLADE Scale Copyright 2010 American Weigh Scales Inc All rights reserved Rev 3 1 US Patent D525 891...

Page 2: ...ation and adjustment if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using y...

Page 3: ...re using the scale for the first time check the battery compartment and remove any plastic insulation strips or packaging that may be in place to prevent battery drainage during shipping storage 2 If...

Page 4: ...will default to the last unit used BLADE 100 Units g gram oz ounce ct carat gn grain BLADE 650 BLADE 1KG Units g gram oz ounce ozt troy ounce dwt pennyweight VI Using the Tare Feature You can use thi...

Page 5: ...ration 1 Place the required calibration weight on the scale and note the displayed value 2 If the displayed value is not within acceptable tolerance see chart below adjustment is required Model Accept...

Page 6: ...e your workspace is stable and free of any air currents or vibration Problem Displayed weight is inaccurate Solution Perform a calibration adjustment X Error Codes 1 OUTZ Zero range has shifted Perfor...

Page 7: ...y 650g 1000g Readability 0 1g 0 1g Min Weight 0 2g 0 2g Power 2 x AAA Dimensions 0 8 x 3 8 x 3 8 Platform Dimensions 2 7 x 2 7 XII Capacity Readability Charts BL 100 100g x 0 01g 3 528oz x 0 001oz 500...

Reviews: