Technical Data and Sensor Accuracy
Table 3: Measuring uncertainties without uncertainty of the calibration gas. The uncertainty
wasdetermined at the calibration point. Values that have been conservatively extrapolated to
themeasuring range end value are indicated in brackets (worst case)
Measured
component
physical
unit
Calibration
point (measur-
ing range end
value)
Standard un-
certainty u2
at delivery
+/– 10°C
Standard un-
certainty u3
after 1 year
+/– 10°C
Standard uncer-
tainty u4 right
after calibration
+/– 0°C
97 (100)
0,1 (0,11)
1,5 (1,6)
0,03 (0,03)
55 (100)
0,1 (0,19)
1 (1,8)
0,03 (0,05)
Methane
Vol.-%
CH
4
2,5 (100)
0,1 (4)
0,4 (1,8)
0,03 (1,2)
45 (100)
0,11 (0,24)
1,3 (2,9)
0,04 (0,09)
Carbon
dioxide
Vol.-%
CO
2
15 (100)
0,1 (0,7)
1,2 (2,9)
0,03 (0,2)
1 (25)
0,05 (1,1)
0,11 (2,7)
0,025 (0,6)
Oxygen
param.
Vol.-% O
2
21 (25)
0,8 (0,9)
1,3 (1,5)
0,1 (0,12)
10 (20)
0,6 (1,2)
1 (17)
0,5 (1)
170 (200)
6 (7)
15 (17)
2,5 (2,9)
170 (500)
8 (24)
16 (50)
6 (18)
1000 (1500)
30 (50)
90 (130)
10 (15)
1000 (3000)
30 (90)
90 (260)
10 (30)
1000 (5000)
30 (170)
90 (400)
15 (80)
Hydrogen
sulphide
Vol.-ppm
H
2
S
1000 (10000)
40 (400)
90 (900)
20 (200)
1000 (1000)
30 (30)
90 (90)
10 (10)
Vol.-ppm
H
2
2000 (5000)
60 (160)
170 (400)
20 (50)
3 (5)
0,1 (0,17)
0,26 (0,4)
0,05 (0,08)
15 (25)
0,5 (0,8)
1,3 (2,2)
0,2 (0,3)
Hydro-
gen elec-
tro-chem.
Vol.-% H
2
100 (100)
3 (3)
9 (9)
0,7 (0,7)
100 (100)
0,5 (0,5)
1,1 (1,1)
0,2 (0,2)
95 (100)
0,9 (1)
1,4 (1,4)
0,8 (0,8)
Hydrogen
TCD
Vol.-% H
2
33 (100)
1,3 (4)
1,6 (5)
0,8 (2,4)
38