INTRODUCTION
The AVD series of monocycle generators allows the generation of high-speed
monocycle waveforms, with peak-to-peak amplitudes of > 20 V into 50Ω loads. The
amplitude of the output waveform is not adjustable.
The monocycle center frequency is fixed. The center frequency is equal to XXXX MHz,
where the model number is AVD-XXXX.
See the online application note, "TB13 - How Is the Center Frequency of a Monocycle
Waveform Measured?" at http://www.avtechpulse.com/appnote/techbrief13/ for
information about center frequency measurements.
The AVD can operate at repetition rates of up to 1 MHz. The AVD must be triggered by
an external TTL pulse (> 50 ns) applied to the “IN” connector.
The output is designed to drive 50Ω loads. (A 50Ω load is required for proper
operation.) The output is AC-coupled.
This instrument is intended for use in research, development, test and calibration
laboratories by qualified personnel.
-F3 OPTION
Units with the -F3 option can be triggered at repetition rates of up to 3 MHz, instead of
the standard 1 MHz limit.
The maximum output amplitude decreases to 15V, approximately, for operation above 1
MHz.