Puro Range Water Purification System Operation Manual
Avidity Science, LLC
8
Operation
Warnings Messages
Below is a list of warning messages and solutions.
Description
Error
Number
Solution
Water leak (system shuts down)
60
Fix the leak. Dry the leak sensor located on the bottom of
the cabinet.
Contact Technical Support as necessary.
Permeate (when RO pump is ON)
40
Contact Technical Support.
Loop Temperature
43
Follow on-screen instructions.
Tank Empty
50
Follow on-screen instructions.
Tank Low
51
Follow on-screen instructions.
Permeate Sensor Fault: LOW
64
Follow on-screen instructions.
Loop Conductivity Sensor Fault
63
Follow on-screen instructions.
Temperature Sensor: HIGH
64
Allow system to flush.
Temperature Sensor: LOW
65
Check incoming water temperature.
Level Sensor Fault:
HIGH
67
Contact Technical Support.
Level Sensor Fault: LOW
68
Contact Technical Support.
Tank UV Lamp Run Time Expired
5
Contact Technical Support
Pretreatment Module TC001 time expired
1
Replace.
Inlet Conductivity: HIGH
24
Check conductivity in Feed Water.
Pressure Sensor Fault: HIGH
69
Contact Technical Support.
Pressure Sensor Fault: LOW
70
Contact Technical Support.
Pressure: HIGH
71
Follow on-screen instructions.
Contact Technical Support.