QC-3265 Checkweigher User’s Manual

To use the QC-3265 as a checkweigher for gross weight, follow the direc-
tions given below:

1.

Zero the scale by pressing the

ZERO 

key or flip the optional

toggle switch down. . .

Display shows 0 weight.

2.

Place a test weight on the scale
equal to the gross weight of the
package. . .

The gross weight is displayed.

3.

Press the 

TARGET 

key or flip

the optional toggle switch up. . .

The display shows 0 weight and the

ACCEPT 

LED illuminates.

4.

Remove the test weight and
place a package on the scale. . .

The display will show the weight
deviation from the target weight and
the 

OVER

ACCEPT

, or 

UNDER

LED will illuminate accordingly.

IMPORTANT NOTE:

If you use the 

ZERO 

key to tare a container weight

and use the checkweigher in net weighing mode, the unit will track zero
whenever there is an empty container on the scale. If the scale is left idle
while empty, zero drift may occur.

1.

Press the 

>

 key. . .

OVER 

is displayed.

2.

Press the 

v

 key. ..

o   0.5 

is an example of what might

be displayed. 

stands for over

tolerance and the value is the
current over tolerance. The actual
value depends on the unit of mea-
sure and division size. The over LED
is lit.

3.

Press the 

key to decrease the

value and the 

key to increase

the tolerance value. Press the 

v

key to accept the displayed value.

OVER

 is displayed.

4.

Press the 

>

 key. . .

UNDER 

is displayed.

5.

Press the 

v

 key. . .

u   0.5 

is an example of what might

be displayed. 

stands for under

tolerance and the value is the
current under tolerance. The under
LED is lit.

6.

Repeat step 3 and press the

QUICK CHECK

 key to return to

the checkweigh or display mode.

Gross Weight
Checkweighing

Changing the
Acceptance Tolerance

Over and Under tolerance can
be set from the User menu or
the Setup menu (see Service
Manual) but not both. The
choice where the tolerances
appear is made in the Setup
menu. The tolerances will not
appear in the User menu if they
appear in the Setup menu and
vice versa.

If you change the tolerance but
decide you would rather keep
the original value, press the 

^

key in step 2 to return to OVER
without accepting the displayed
value.

Target weight and the toler-
ances are saved in nonvolatile
memory. They are not lost
when the unit is powered
down.

In gross mode, the indicator
will always represent negative
weight with a leading minus
sign. If the weight becomes too
large, positive or negative,
overscores or underscores will
be displayed while the condi-
tion persists.

Summary of Contents for QC-3265

Page 1: ...QC 3265 Checkweigher User s Manual...

Page 2: ...3265_U P65 PN 29600 0011J e1 Printed in USA Risk of electrical shock Do not remove cover No user service able parts inside Refer servicing to qualified service personnel Weigh Tronix reserves the righ...

Page 3: ...and Operation 9 Installation 9 Explanation of Modes of Operation 9 Operation as a Scale 9 Gross Weight Checkweighing 10 Net Weight Checkweighing 10 Changing the Acceptance Tolerance 11 User Menu 11 Co...

Page 4: ...V C Span 5 0 ppm C 10 ppm C For scale base 3000 divisions 0 005 of capacity maximum For base 0 01 0 005 of capacity maximum For base 0 01 0 005 of capacity maximum For base 0 01 All calibration and p...

Page 5: ...00 x 02 3 x 0005 3000 x 5 12 x 005 200 x 0 1 6 x 002 6000 x 2 12 x 002 200 x 0 05 6 x 001 6000 x 1 30 x 01 480 x 0 2 15 x 005 15000 x 5 30 x 005 480 x 0 1 15 x 002 15000 x 2 60 x 02 960 x 0 5 30 x 01...

Page 6: ...6 QC 3265 Checkweigher User s Manual This page left intentionally blank...

Page 7: ...Weigh Tronix distributor Figure 1 shows the front panel The panel consists of the following a six digit liquid crystal or LED display five keys QUICK CHECK TARGET UNITS ZERO Right arrow key a center...

Page 8: ...of measure is enabled Use this key to move to the left in the menus Use the ZERO down arrow key to zero the scale while in either display mode In the A D test mode this key sets the offset to zero Us...

Page 9: ...rs up in weight display mode if there is no target weight in memory Use the checkweigh mode for checkweighing operations and the use of the OVER ACCEPT and UNDER LEDs If there is a target weight in me...

Page 10: ...mea sure and division size The over LED is lit 3 Press the key to decrease the value and the key to increase the tolerance value Press the v key to accept the displayed value OVER is displayed 4 Pres...

Page 11: ...ipped with an LED weight display ABOUT Press the v key with ABOUT displayed and QC3265 will be displayed Press the key to step through the EPROM part number dash number revision letter and then back t...

Page 12: ...ulator A to D Shows the output of the A to D converter with a sensitivity of 20 000 counts per millivolt per volt SERIAL Lets you perform READY BUSY and LOOP NOLOOP tests to check the serial port READ...

Page 13: ...e is detected This appears when voltage level reaches 10 5 volts and alternates with the normal display The unit will shut itself off at 9 2 volts Only available on battery powered units Communication...

Page 14: ...Installation for USDA Approved Applications Figure 4 Plastic knob installation Below are the instructions and illustration for installing the plastic knobs which are necessary for USDA approved insta...

Page 15: ...irmont MN 56031 USA Telephone 507 238 4461 Facsimile 507 238 4195 e mail industrial weigh tronix com www wtxweb com 217 Brunswick Blvd Pointe Claire QC H9R 4R7 Canada Telephone 514 695 0380 Facsimile...

Reviews: