Minute ventilation
High
1 ~ 99 L
Low
0 ~ 98 L
Respiratory rate
High
1 ~ 100bpm
Low
OFF,1 ~ 99 bpm
FiO2(optional)
High
19 ~ 100%, OFF
Low
18 ~ 99%
Airway pressure
High
6 ~ 99 cmH2O
Low
0 ~ 98 cmH2O
SpO2(optional)
High
50 ~ 100 %
Low
49 ~ 99 %
Pulse(optional)
High
31 ~ 250 bpm
Low
30 ~ 249 bpm
ETCO2 (optional)
High
0.1 to 13.3%;
Low
0 to 13.2%
Apnea alarm
Setting time is 10 ~ 60 s, in increments of 1 s.
Drive gas failure alarm
Drive gas pressure is less than 280 kPa.
AC power failure alarm
The mains breaks down or the power cord
disconnects
Low battery alarm
Alarm battery time is 20 min.
Battery discharged alarm
Alarm battery time is 10 min.
O2 sensor failure
FIO2 <15% Vol.
Silent Alarm
≤ 120 s
A.8 Anesthetic Vaporizer (optional)
Anesthetic vaporizer (for details, refer to the vaporizer Instructions for Use)
Vaporizer position
double vaporizer positions
Mounting mode
with interlocking function
Applicable standard
ISO 80601-2-13