Chapter 3 Installation and Connections
Warning
◆
To ensure that pipeline gas supply hose and breathing circuit components are not toxic, and
will not cause allergic reactions to the patient and does not react with the anesthetic
substances.
◆
Continuous use of desiccated soda lime may endanger patient safety. Adequate precautions
should be taken to ensure that the soda lime in the absorber canister does not become
desiccated. Turn off all gases when finished using the system.
◆
When using electrical surgical equipment, keep wires away from the breathing circuit, oxygen
sensors, flow sensors and other components on the Anesthetic workstation and ensure that
the Anesthetic workstation manual ventilation mode can be used at any time in order to avoid
interference with electrical surgical equipment. In the meantime, please note that all
monitoring and life support equipment are operation ready.
◆
Never use antistatic or conductive masks with the breathing tubes. They can cause burns if
they are used near high frequency electrosurgical equipment.
◆
This equipment must be installed by the company authorized engineer.
◆
This Anesthetic machine has waste gas exhaust ports. The operator of the machine should
pay attention to the disposal of the residual breathing gas scavenged.
◆
The Anesthetic workstation and power usage environment must meet environmental
specifications B.2 B.3 specifications inside power requirements.
◆
Before Installing and using the Anesthetic workstation, all packing materials should be
removed completely.
3.1 Install CO2 Circle Absorber
Caution
⚫
Pay attention to the disposal of the breathing system after equipment use, the detection of the
soda lime in the canister and the Anesthetic agent in the vaporizer to ensure the normal
operation of the equipment.
3.1.1 install the arm
Install the arm on the aluminum piece of the frame and use the black knob to tighten it.
3.1.2 install the CO2 Circle absorber
1.Put the metal body on the arm, and use black knob to tighten.
2.Use 4 screws to secure the aluminum board on the arm, then insert the bellow on the aluminum
board.
3.Install the bag arm on the metal body.
4.Install the pressure gauge on the metal body.