5.Install the soda lime chamber on the down of metal body Insert the soda lime chamber.
3.2 Install the Oxygen Sensor
Warning
◆
Before installing the O
2
sensor, check that the seal on the sensor is in good condition. Fit a
new sensor if the seal is missing or damaged.
◆
Oxygen sensor must be tightened to avoid breathing system leak.
1. Install the three way connector on the expiration connector and insert the O2 sensor on it.
2. Insert the sensor cable into the O2 sensor.
3. Insert the other end of the cable into the O
2
sensor connector marked O
2
% on the rear of the
Anesthetic machine frame.
3.3 Install the breath circuit
Caution
⚫
When installing the breathing tube, hold the tube connector at both ends of the tube to prevent
damage of the tube.
1. The two breathing hoses are connected to the breathing system with expiratory and inspiratory
connectors.
3.4 Install flow sensor
1.Insert the Flow Sensor between the Y-piece of the breath circuit and the patient connection as
FIGURE 3-1.
The blue tube of flow sensor is closest to the patient.
Connect the blue and colorless tubes to the Flow Sensor connectors in the back of anesthesia
machine. The blue tube goes to the connector with blue mark. The colorless tube goes to the
connector with colorless mark.
FIGURE 3-1
Installing the Flow Sensor
NOTE:
To prevent inaccurate Flow Sensor readings, make sure the Flow Sensor is correctly installed: