12.4 Enabling or Disabling non-volatile Dose Storage
This function allows to view and modify dose storage mode. The current setting is displayed as follows:
Stor
(= internal dose is stored, factory setting),
or
noSt (= no storage of internal dose, after switching on or disconnecting a probe the dose
is always zero).
The messages »Stor« and »no« flash to indicate that they are subject to modification. If you wish to
modify the setting, press the signal key accordingly. Store the setting by pressing and holding down the
on/off key and then pressing the function key.
If you do not wish to modify the setting, press the function key to move to the next programming function
(see next section).
12.5 Resetting to Factory Setting
This function allows to reset all programmable data to factory setting. With some 6150AD models it also
allows to select the unit (Sv or R). The display looks as follows:
Sv
(= unit is Sv and Sv/h, factory setting),
rSEt
or
R
(= unit is R and R/h).
rSEt
The unit »Sv« or »R« flashes to indicate that it is subject to modification. If you wish to modify the unit,
press the signal key accordingly. Store the new unit by pressing and holding down the on/off key and then
pressing the function key. Note that this will also reset all data to factory settings and clear the internal
dose. It is not possible to change the unit without resetting all other data. That is no considerable
restriction, because selecting the unit is subject to national regulations and is usually required only once.
The option to select the unit is not available for H*(10) models (/H and /E models), because H*(10) is
always measured in Sv units. With such models this function only serves to reset to factory setting; the
display then only shows »rSEt« leaving the unit blank.
The factory settings are summarized below:
All 14 individual alarm thresholds are set to »OFF«.
The post-alarm increment is set to 2%.
The storage mode for the internal dose is set to »Stor«.
The internal dose is reset to zero.
Note that, when resetting, the internal dose is cleared although the storage mode »Stor« is set
simultaneously. The »Stor« mode thus applies to uses following the reset.