6.5 Dose
Internal tubes and Sv-probes use three digits with floating decimal point and variable unit. With the
6150AD1/2 the smallest step is 1 µSv, and with the 6150AD3/4/5/6 the smallest step is 0.01 µSv.
With pulse probes, the smallest step is always 1 (one pulse). The 6150AD1/2 indicates integer pulse
counts up 9999, whereas the 6150AD3/4/5/6 uses three digits with floating decimal point and variable
unit.
internal tubes and Sv-probes
pulse probes
6150AD1/2
6150AD3/4/5/6
6150AD1/2
6150AD3/4/5/6
0.00 - 9.99 µSv
10.0 - 99.9 µSv
0 - 999 µSv
100 - 999 µSv
0 - 999
1.00 - 9.99 mSv
10.0 - 99.9 mSv
100 - 999 mSv
1.00 - 9.99 Sv
0 - 9999
1.00 - 9.99 k
10.0 - 99.9 k
100 - 999 k
The table shows the 6150AD’s entire range. The upper limit of the range will automatically be restricted
to the dose range of the corresponding detector (internal tube or probe).
When exceeding the highest range, the flashing full scale value (»9...9«) will indicate overrange.
6.6 Dose Alarm Threshold (6150AD3/4/5/6 only)
This section only applies to the 6150AD3/4/5/6, because the 6150AD1/2 does not provide dose alarm
thresholds. The smallest step is always 1 µSv or 1 (one pulse), respectively.
internal tubes and Sv-probes
pulse probes
0 - 999 µSv
1.00 - 9.99 mSv
10.0 - 99.9 mSv
100 - 999 mSv
1.00 - 9.99 Sv
0 - 999
1.00 - 9.99 k
10.0 - 99.9 k
100 - 999 k
The table shows the 6150AD’s entire range. The upper limit of the range will automatically be restricted
to the dose range of the corresponding detector (internal tube or probe).
Page 26
Operating Manual 6150AD
03/2004