For pulse probes there are six pulse rate ranges that will also be selected automatically:
pulse rate ranges:
range no.
1
2
3
4
5
6
analog scale start
analog scale centre
analog scale end
0.1
1
10
1
10
100
10
100
1000
0.1
1
10
1
10
100
10
100
1000
unit
S
-1
S
-1
S
-1
kS
-1
kS
-1
kS
-1
digital display format:
probes
6150AD-17,
6150AD-19
0.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
probe 6150AD-0
0.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
When exceeding the highest range, the flashing full scale value (»9...9«) will indicate overrange.
Remark on the probe 6150AD-0: this is a general purpose probe type, the pulse rate of which is indicated
directly without applying any calibration factor. We created this probe type for future developments of
pulse probes that will connect to any 6150AD model. Because one cannot foresee the range of a future
pulse probe, the 6150AD-0 range goes up to 999 kS
-1
, that is approx. 1 MHz, which is about the
maximum count capacity of the 6150AD. That high range is a reserve for the future not required by any
of the probes currently available.
6.2 Dose Rate Mean Value
Ranges and formats are equal for all 6150AD models.
internal tubes and Sv-probes
pulse probes
0.000 - 9.999 µSv/h
10.00 - 99.99 µSv/h
100.0 - 999.9 µSv/h
1.000 - 9.999 mSv/h
0.000 - 9.999 S
-1
10.00 - 99.99 S
-1
100.0 - 999.9 S
-1
1000 - 9999 S
-1
The range is the same for all detectors (internal tubes and probes). However, the Scintillator Probe
6150AD-b cannot use the full range because its dose rate range ends at 99.9 µSv/h. You may observe
mean values greater than 99.9 µSv/h with the 6150AD-b, but such mean values are not reliable because
they are out of range.
When exceeding the highest range, the flashing full scale value (»9...9«) will indicate overrange.