CALIBRATION PROCEDURE
The Mini Plus must be re-calibrated every two years. The reason for this is to maintain
creditability in tests and acceptance of data according to international standards. Each
time the unit is turned on; the LCD will show the date that the re-calibration is due. At the
time the unit is due for re-calibration, the LCD screen will show:
“Calibration Due!
Please contact your vendor for re-
calibration”
, and further testing will be disabled.
There are two ways of calibrating the Brake Meter - either return the AutoStop Mini Plus
to your local distributor or by returning the Brake Meter to Auto Test Products (See
Section 6.1).
6.1 Return of AutoStop Mini Plus for Calibration
6.1.1 Packaging
Please remember that you are shipping an electronic instrument. Bubble pack or foam
should surround the Mini Plus and should be inserted into a sturdy cardboard box.
6.1.2 Shipping
Labelling
-
A label noting “Electronic Device - Fragile” should be placed on the box.
Freight Carrier
-
Container should be sent “Freight Prepaid”. Auto Test Products has no
preference on freight carriers. However, we prefer that companies such as TNT, IPEC,
UPS, or Federal Express forward units, if a prompt delivery is required.
Ship to
: The Service Department, Auto Test Products Pty Ltd, 279 Normanby Road, Port
Melbourne, VIC 3207, Australia. Phone: (+61 3) 9647 9797, or to your nearest service
distributor.
6.1.3 Documentation
A copy of the purchase order for re-calibration and a packing slip must be included with
the shipment. It should note the Purchase Order Number, Date, Serial Number and the
address where the unit should be returned.
7
TROUBLESHOOTING
Problem
Potential Cause and Action Necessary
Mini Plus will not turn on
Internal electronics have failed - return unit for
servicing.
Mini Plus not recording keypad
input
Keypad damaged- return unit for servicing.
Mini Plus turns itself off during
use
Battery flat-battery needs to be recharged.
Mini Plus won’t run tests
Calibration due - return unit for calibration.
Mini Plus displays
Clock failed
The in-built clock has failed. It might be an electronic
fault, or the unit needs to be recalibrated.
Mini Plus displays
Error
The test was not carried out correctly; repeat the test.