Augmedics Ltd • xvision-Spine User Manual
Version 7
Test
Frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
Power (W)
Distance
(m)
Immunity
Test Level
(V/m)
Compliance
Level (V/m)
5240
5 100 – 5 800
WLAN 802.11 a/n
Pulse
modulation
b
217
Hz
0.2
0.3
9
9
5500
5785
Note:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50% duty cycle square wave signal.
c
As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because, while it does not represent actual
modulation, it would be the worst case.
Test Specifications for Enclosure Port Immunity to RF Wireless Communications Equipment
(Headset and Optional Tablet) (continued)
2.7. Method of Sterilization or Disinfection
The sterile components of the
XVS
system are supplied gamma radiation sterilized and are
intended for single use only.
on page 8-1 and
(page 8-3) for
information on cleaning and sterilization of reusable components of the
XVS
system.
The other parts of the
XVS
system should not be sterilized.
2.8. Manufacturer Responsibility
Augmedics Ltd is responsible for the safety, reliability and performance of the
XVS
system
only if:
• Assembly, operations, extensions, modifications, service and repairs are carried out by
authorized Augmedics personnel
• The
XVS
system is used in accordance with this User Manual and all applicable safety
regulations
2.9. General Notes, Cautions, and Warnings
Warning:
The system is not suitable for use in the presence of an anesthetic flammable
mixture with air or oxygen or nitrous oxide.
2-10