Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
3V, 6V
3Vrms, 6V
Portable and mobile RF communications
equipment should be used no closer to any part
of the
XVS
, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended Separation Distance
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range. D Interference may occur in the
vicinity of equipment marked with the following
symbol:
Radiated RF IEC 61000-
4-3
3V/m
3V/m
3V from 0.15 to
80MHz;
6V from 0.15 to
80MHz and 80% AM at
1kHz
3V from 0.15 to
80MHz;
6V from 0.15 to
80MHz and 80% AM at
1kHz
10V/m from 80MHz to
2.7GHz
10V/m from 80MHz to
2.7GHz
Declaration: Electromagnetic Immunity (Optional Tablet and Router) (continued)
2-8