2. Safety Information
2.5. Cybersecurity User Instructions
likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
2.5. Cybersecurity User Instructions
The
XVS S
OFTWARE
requires a wireless router configured as described in
Augmedics recommends that users enable a firewall on the hospital PC and configure the
firewall so that only the
XVISION
-S
PINE
network is enabled.
Augmedics Ltd uses industry-standard instructions to protect the
XVISION
-S
PINE
system. All
events are logged to the Windows Security Event log.
In addition, Augmedics maintains regular security patches for operating systems, applications,
and modules used.
Only network ports (inbound) 24967, 24969 to 24975 (TCP), 104 (TCP) and the port configured
for DICOM access (as defined by the installed scanner) should be opened for send/receive. All
other ports should be disabled. See
Configuring the Scanner Imaging Protocol
on page 5-6 for
information about DICOM access.
Hospital PC Security
It is the responsibility of the authorized user to ensure that the hospital PC on which the
XVISION
-S
PINE
application is installed is not left unattended, unlocked, or otherwise unsecured
when not in use, to ensure that non-authorized medical, professional, or otherwise
unapproved personnel are not exposed to, or gain access to, ePHI. The same applies to the
pre-configured Augmedics PC.
Anti-Virus and Anti-Malware
Augmedics recommends that users install an anti-virus and an anti-malware application on
the hospital PC and run regular scans on the PC to monitor, detect and prevent viruses or
malware on their PC. A pre-configured Augmedics PC includes antivirus and antimalware
utilities.
User Responsibility
Use of the
XVISION
-S
PINE
application and the PC on which it runs is the responsibility of the
end user. To prevent potential unauthorized access, the user should not leave the PC
unattended or in the possession of a non-authorized user. Augmedics uses industry-standard
instructions to protect the
XVS S
OFTWARE
system. All events are logged to the Windows
Security Event log.
Caution:
If XVS Software is installed by a third-party (the hospital) it is the responsibility of
the third party to secure the network.
2-3