![Audio Precision AP 517 Installation Instructions Manual Download Page 82](http://html1.mh-extra.com/html/audio-precision/ap-517/ap-517_installation-instructions-manual_3008579082.webp)
80
APx PDM I/O Module for B Series: Specifications
PDM
Clock Jitter (Advanced Master Clock required)
Jitter Measurement
Range
0 to 650 ns
Detection
Peak, RMS or Average
Bandwidth
Low Limit
50 Hz or 700 Hz
High Limit
Variable in 0.1 kHz steps,
Butterworth or Elliptic response
1 kHz
150 kHz
Accuracy (1 kHz)
“Average” detection
±(1% + 300 ps)
Flatness
1
100 Hz to 100 kHz
±0.2 dB
Residual Jitter
1
700 Hz to 100 kHz BW
≤
600 ps
Jitter Spectrum
1
Spurious products
are typically
–40 dBc (below jitter
signal) or
–60 dBUI,
whichever is larger.
2
PDM Input Jitter Tolerance
Sine wave jitter, bit clock rates
from 128kHz to 24.576 MHz.
3.5 UI,
(subject to
1591 ns max
jitter limit)
Induced Jitter
Waveforms
Sine, Square, Noise
Signals Affected
Bit Clk and Data
Parameter
Symbol
Test Conditions
Min
Typ
Max
Unit
Summary of Contents for AP 517
Page 1: ......
Page 2: ......
Page 6: ......
Page 14: ...S curit 12 APx517 B Series Acoustic Analyzer S curit...
Page 18: ...Seguridad 16 APx517 B Series Acoustic Analyzer Seguridad...
Page 24: ...Installation 22 APx517 B Series Acoustic Analyzer Installation...
Page 30: ...Installation Fr 28 APx517 B Series Acoustic Analyzer Installation Fr...
Page 36: ...34 Abbreviations Terms and Symbols...
Page 52: ...50 APx517 B Series Acoustic Analyzer Specifications...
Page 84: ...82 APx PDM I O Module for B Series Specifications PDM...
Page 91: ......
Page 92: ...Audio Precision Inc 5750 SW Arctic Drive Beaverton Oregon 97005 800 231 7350...