![Audio Precision AP 517 Installation Instructions Manual Download Page 47](http://html1.mh-extra.com/html/audio-precision/ap-517/ap-517_installation-instructions-manual_3008579047.webp)
Characteristic
Specifications
Supplemental Information
APx517 B Series Acoustic Analyzer: Specifications
45
IMD Measurement
Test Signal Compatibility
SMPTE & MOD
Any combination of 40 Hz - 1 kHz
(LF)
and 240 Hz-60 kHz (HF) tones,
mixed
in any ratio from 1:1 to 10:1 (LF:HF)
LF tone must be
≤
1/6 * HF tone
DFD & CCIF
Any two-tone combination with mean
frequency of 250 Hz-60 kHz and a
difference frequency of 80 Hz-2.0
kHz
Fmean = (F1 + F2)/2; Fdiff = |F2 - F1|
Fmean must be
≥
6*Fdiff
IMD Measured
SMPTE
Amplitude modulation of HF tone
Measurement BW is
≈
40-750 Hz
MOD
d2, d3, d2+d3, or d2+d3+d4+d5
Use "d2+d3" for measurements per IEC-
60268
DFD
d2, d3, d2+d3, or d2+d3+d4+d5
Use "d2+d3" for measurements per IEC-
60268
CCIF
d2 only
"CCIF" is an archaic form of DFD that
measures only the d2 product. CCIF
uses a different 0 dB reference giving
readings 2x higher than DFD
Measurement Range
0 to 20%
Accuracy
±0.5 dB
Residual IMD {notes 1,2,3}
SMPTE & MOD
≤
-90 dB [0.0032%], 4:1 mix ratio
DFD
≤
-100 dB [0.0010%]
Summary of Contents for AP 517
Page 1: ......
Page 2: ......
Page 6: ......
Page 14: ...S curit 12 APx517 B Series Acoustic Analyzer S curit...
Page 18: ...Seguridad 16 APx517 B Series Acoustic Analyzer Seguridad...
Page 24: ...Installation 22 APx517 B Series Acoustic Analyzer Installation...
Page 30: ...Installation Fr 28 APx517 B Series Acoustic Analyzer Installation Fr...
Page 36: ...34 Abbreviations Terms and Symbols...
Page 52: ...50 APx517 B Series Acoustic Analyzer Specifications...
Page 84: ...82 APx PDM I O Module for B Series Specifications PDM...
Page 91: ......
Page 92: ...Audio Precision Inc 5750 SW Arctic Drive Beaverton Oregon 97005 800 231 7350...