![Audio Precision AP 517 Installation Instructions Manual Download Page 48](http://html1.mh-extra.com/html/audio-precision/ap-517/ap-517_installation-instructions-manual_3008579048.webp)
Characteristic
Specifications
Supplemental Information
46
APx517 B Series Acoustic Analyzer: Specifications
Frequency Measurement
Range
< 5 Hz to 90 kHz
Accuracy
±(0.0003% + 100
μ
Hz)
Vin must be
≥
5 mV
Resolution
6 digits
Phase Measurement
Ranges
-90 to +270, ±180, or 0 to 360 deg
Accuracy
±0.2 deg, 5 Hz to 5 kHz;
±0.8 deg, 5 kHz to 20 kHz;
±2.0 deg, 20 kHz to 50 kHz
Vin must be
≥
5 mV with DC coupling,
both channels. Accuracy degrades
below 50 Hz with AC coupling
Resolution
0.001 deg
DC Voltage Measurement
Input Ranges
0.25 V to 25 V, 10 dB steps
±40 Vdc maximum in 25 V range.
Accuracy
250 mV and 800 mV ranges
±(0.7% r 1 mV)
2.5 V-8 V ranges
±(0.7% r 0.1% range)
Normal Mode Rejection
Typically > 90 dB, 20 Hz to 20 kHz
Summary of Contents for AP 517
Page 1: ......
Page 2: ......
Page 6: ......
Page 14: ...S curit 12 APx517 B Series Acoustic Analyzer S curit...
Page 18: ...Seguridad 16 APx517 B Series Acoustic Analyzer Seguridad...
Page 24: ...Installation 22 APx517 B Series Acoustic Analyzer Installation...
Page 30: ...Installation Fr 28 APx517 B Series Acoustic Analyzer Installation Fr...
Page 36: ...34 Abbreviations Terms and Symbols...
Page 52: ...50 APx517 B Series Acoustic Analyzer Specifications...
Page 84: ...82 APx PDM I O Module for B Series Specifications PDM...
Page 91: ......
Page 92: ...Audio Precision Inc 5750 SW Arctic Drive Beaverton Oregon 97005 800 231 7350...