![Audio Precision AP 517 Installation Instructions Manual Download Page 39](http://html1.mh-extra.com/html/audio-precision/ap-517/ap-517_installation-instructions-manual_3008579039.webp)
Characteristic
Specifications
Supplemental Information
APx517 B Series Acoustic Analyzer: Specifications
37
Non-Harmonic Content
Typically < -100 dB, 20 Hz to 20 kHz
DC Offset Range
±3.00 Vdc unbalanced
DC offset limits maximum ac signal
Residual DC Offset
+/-5 mV
typically < 650 uV
Noise Characteristics
Shape
White or Pink (5 Hz to 86.4 kHz),
IEC 60268-1, or BS EN 50332-1
Amplitude Range
0 to 48.08 Vpp
Amplitude calibration is approximate
Residual Noise {Note 2}
20 Hz to 20 kHz BW
≤
30 uVrms [-88.8 dBu]
IMD Test Signals
SMPTE & MOD
Lower Frequency (LF)
40 Hz to 1.00 kHz
LF tone must be
≤
1/6 * HF tone
SMPTE Upper Frequency
(HF)
2.00 kHz to 60.00 kHz
MOD Upper Frequency (HF)
240 Hz to 60.00 kHz
Mix Ratio (LF:HF)
10:1, 4:1, or 1:1
Amplitude Range
0 to 48.08 Vpp
Amplitude Accuracy
±0.06 dB [±0.70%]
Residual IMD {notes 1,2,3}
≤
0.020% [-74 dB], 4:1 mix ratio
DFD & CCIF
Difference Frequency (Fdiff)
80 Hz to 2.00 kHz
Fdiff = |F2 - F1|
Fmean must be
≥
6*Fdiff
Mean Frequency (Fmean)
250 Hz to 60.00 kHz
Fmean = (F1 + F2)/2
Summary of Contents for AP 517
Page 1: ......
Page 2: ......
Page 6: ......
Page 14: ...S curit 12 APx517 B Series Acoustic Analyzer S curit...
Page 18: ...Seguridad 16 APx517 B Series Acoustic Analyzer Seguridad...
Page 24: ...Installation 22 APx517 B Series Acoustic Analyzer Installation...
Page 30: ...Installation Fr 28 APx517 B Series Acoustic Analyzer Installation Fr...
Page 36: ...34 Abbreviations Terms and Symbols...
Page 52: ...50 APx517 B Series Acoustic Analyzer Specifications...
Page 84: ...82 APx PDM I O Module for B Series Specifications PDM...
Page 91: ......
Page 92: ...Audio Precision Inc 5750 SW Arctic Drive Beaverton Oregon 97005 800 231 7350...