background image

22

Electrical Safety and Compatibility

Appendix

04

Symbol

Description

Symbol

Description

Symbol

Description

Symbol

Description

Manufacturer

MR unsafe

Ingress Protection

Caution

Date of manufacture 

and country of 

manufacture

(made in Korea)

Atmospheric pressure

limitation

Type BF Applied part

Separate collection

for EEE

Unique Device

Identifier

Humidity limitation

Direct Current

Medical Device

Model number

Temperature limit

Prescription Only

Professional Use Only

Not available without a 

doctor’s prescription

Do not re-use

Serial number

Keep away from

sunlight

Use by It can be used

up to the date stated

by the manufacturer

Keep dry

Refer to instruction 

manual

/ booklet

Do not use if package is

damaged and consult 

instructions for use

CE Marking of

Conformity

Authorized representative

in the European 

Community

/ European Union

Non-ionizing radiation

Symbols

Appendix

04

KR

23

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF Emissions

CISPR 11

Group 1

The device uses RF energy only for its internal function. Therefore, its RF emissions are

very low and are not likely to cause any interference in nearby electronic equipment.

RF Emissions

CISPR 11

Class B

The device is suitable for use in all establishments, including domestic establishments and those directly

connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Immunity Test

IEC 60601 Test and Compliance Level

Electromagnetic Environment - Guidance

Electrostatic Discharge (ESD) 

IEC 61000-4-2

± 8 kV/Contact 

± 2, ± 4, ± 8, ± 15 kV/Air

Floors should be wood, concrete, or ceramic tile. If floors are covered

with synthetic material, the relative humidity should be at least 30 %.

Power Frequency (50/60 Hz) 

Magnetic Field IEC 61000-4-8

30 A/m

50 Hz or 60 Hz

Power frequency magnetic fields should be at levels characteristic of

a typical location in a typical commercial or hospital environment.

Radiated RF

IEC 61000-4-3

Test Frequency

Level

Portable and mobile RF communications equipment should be used

no closer to any part of the device, including cables, than the

recommended separation distance calculated from the equation

applicable to the frequency of the transmitter.

Recommended separation distance

                                          d = 1.2√P
                                          d = 1.2√P 80 MHz to 800 MHz
                                          d = 2.3√P 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter

in watts (W) according to the transmitter manufacturer and d is

the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by

an electromagnetic site survey, should be less than the

compliance level in each frequency range.

Interference may occur in the vicinity of equipment marked

with the following symbol :

80 MHz - 2,7GHz

80% AM at 1 kHz

10 V/m

385MHz

(Pulse Modulation 18Hz)

27 V/m

450MHz

(FM ±5 kHz deviation 1 kHz sine)

28 V/m

710, 745, 780 MHz

(Pulse Modulation 217Hz)

9 V/m

810, 870, 930 MHz

(Pulse Modulation 18Hz)

28 V/m

1720, 1845, 1970, 2450 MHz

(Pulse Modulation 217Hz)

28 V/m

5240, 5500, 5785 MHz

(Pulse Modulation 217Hz)

9 V/m

30 kHz

(CW) 

8 A/m

134.2 kHz

(Pulse Modulation 21.1 kHz)

65 A/m

13.56 MHz

(Pulse Modulation 50 kHz)

7.5 A/m

Summary of Contents for AT-Patch

Page 1: ...AT Patch User Manual P94A000001A DUG 02 EU Rev 0 02 2022 This product is a disposable medical device Reuse is prohibited...

Page 2: ...attach the device Additional Information 4 7 Caution Clinical information Specifications Electrical Safety and Compatibility Symbols 18 20 21 22 23 3 Introduction 01 Product Components Sub Patch 1EA M...

Page 3: ...least 2 minutes Cleaning the area where the device will be attached with alcohol How to attach the device How to use the device 02 How to use the device 02 5 2 3 4 Turning on the power Press the power...

Page 4: ...at once Reattachment can not be possible Attaching the Sub Patch 7 After attaching the Sub Patch detach the protective tape from the base film slowly Removing protective tape 8 Checking the attachment...

Page 5: ...ps then start using the App Supports Android 5 0 or later iOS 11 0 or later How to install the App 8 Screen with the App AT Note Icon Initial Run Screen How to use the App 03 How to use the App 03 No...

Page 6: ...e Symptom Note entry screen 6 A section displaying the ECG and 3 axis data output through graphs in real time LIVE Display Screen No Description No Description No Description 1 Stop generating real ti...

Page 7: ...the App receives the event information from the device and a dialog box is displayed to ask whether to record the symptoms immediately in a pop up window If the Later is tapped the Symptom Note entry...

Page 8: ...gistered at the App AT Note Upon tapping each revision icon or revision icon area on a single entry of Symptom Note the previously selected values for the items are displayed on the screen Empty value...

Page 9: ...data viewing screen is displayed upon tapping the red rectangular section No Description 1 ECG recording start time and end time 2 Move to ECG recording data viewing screen icon 3 Go Back icon 4 Move...

Page 10: ...could result 11 Cautions should be communicated to the patient 12 All serious accidents related to the medical devices should be reported to the health authorities of the Member State where the manufa...

Page 11: ...th waste disposal regulations may result in environmental pollution NOTE The data stored within the device must be taken care of before disposal Conditions for Use 1 Temperature range 10 45 2 Relative...

Page 12: ...Immunity Immunity Test IEC 60601 Test and Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 8 kV Contact 2 4 8 15 kV Air Floors should be wood concrete or...

Page 13: ...s Co Ltd All Rights Reserved ATsens Co Ltd Point Town 803 11 Gumi ro Bundang gu Seongnam si Gyeonggi do 13637 Republic of Korea Tel 82 70 5220 0220 CMC Medical Devices Drug S L C Horacio Lengo n18 C P...

Reviews: