Contraindications

  1) Do not use if you have current symptoms or medical history of skin cancer, rash, dermatosis, keloids, wounds, etc.

  2) Do not use patients with cardiac pacemakers, cardiac defibrillators or other implantable electrical devices.

  3) Not for use on pregnant women, breastfeeding mothers, infants or children.

Adverse Event 

Physician should advise patients to seek medical attention if either of the following occurs.

  1) Mild rash

  2) Allergic reaction

  3) Skin irritation

  4) Itchiness

Warnings

  1) This product is for single use only. Reuse is prohibited. Reusing it may lead to malfunction and inaccurate results.

  2) Do not attach to any other place than the recommended bodily location of Application.

  3) Only authorized technicians are allowed to repair or disassemble Device.

  4) Make sure to be fully aware of how to use the product, through sufficient training, before using it. 

  5) Avoid use at locations where wireless communication interference may occur (e.g., places where metallic hardware or electronic 

       devices are frequently found).

  6) This product cannot be simultaneously used with a defibrillator. 

  7) Do not expose to strong electromagnetic fields.

  8) The device should not be used adjacent to or stacked with other equipment.

  9) The device may be interfered with by other equipment, even if that other equipment complies with CISPR EMISSIONS requirements.

10) Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no 

       closer than 30 cm (12 inches) to any part of the device. Otherwise, degradation of the performance of this equipment could result. 

11) Cautions should be communicated to the patient.  

12) All serious accidents related to the medical devices should be reported to the health authorities of the Member State where the

       manufacturer, user and patient are located. 

18

Caution

Appendix

04

Precaution

  1) Inappropriate Application and use of the sensor may result in inaccurate measurements, please avoid the below.

     - Excessive movement of the patient.

     - Application outside the recommended bodily locations.

     - To prevent signal abnormalities due to the state of the patient’s skin, sufficient notice of use from physician must be obtained

        before use.
  2) The following persons should consult physician before using the device.

     - Patients with sensitive skin or skin allergy.

     - Patients with wounds on the skin that contact the device.
  3) Take caution to avoid getting liquid in the device, and also avoid the following.

     - Long-term exposure to water such as baths, swimming, etc.
  4) Avoid letting the device come into contact with organic compounds like thinners or benzene.
  5) Beware of strong shocks and vibrations.
  6) Once the product is detached from the body, do not reattach it. 
  7) The smartphone on which the App is run must be one with a test report or a certificate issued by national or accredited offices.
  8) The device needs special precautions regarding EMC and needs to be utilized according to the EMC information provided in the 

       Electrical safety and compatibility section.
  9) Portable and mobile RF communications equipment can affect medical electrical equipment. 
10) If the device stops in the middle, see physician.  
11) Measures to be taken when a cybersecurity issue occurs., 

     - The manufacturer should be consulted if a cybersecurity issue occurs.

       ATsens Co., Ltd. : Tel. +82 70 5220 0220  /  Fax. +82 70 8270 0738  /  E-mail. 

[email protected] 

19

Appendix

04

Summary of Contents for AT-Patch

Page 1: ...AT Patch User Manual P94A000001A DUG 02 EU Rev 0 02 2022 This product is a disposable medical device Reuse is prohibited...

Page 2: ...attach the device Additional Information 4 7 Caution Clinical information Specifications Electrical Safety and Compatibility Symbols 18 20 21 22 23 3 Introduction 01 Product Components Sub Patch 1EA M...

Page 3: ...least 2 minutes Cleaning the area where the device will be attached with alcohol How to attach the device How to use the device 02 How to use the device 02 5 2 3 4 Turning on the power Press the power...

Page 4: ...at once Reattachment can not be possible Attaching the Sub Patch 7 After attaching the Sub Patch detach the protective tape from the base film slowly Removing protective tape 8 Checking the attachment...

Page 5: ...ps then start using the App Supports Android 5 0 or later iOS 11 0 or later How to install the App 8 Screen with the App AT Note Icon Initial Run Screen How to use the App 03 How to use the App 03 No...

Page 6: ...e Symptom Note entry screen 6 A section displaying the ECG and 3 axis data output through graphs in real time LIVE Display Screen No Description No Description No Description 1 Stop generating real ti...

Page 7: ...the App receives the event information from the device and a dialog box is displayed to ask whether to record the symptoms immediately in a pop up window If the Later is tapped the Symptom Note entry...

Page 8: ...gistered at the App AT Note Upon tapping each revision icon or revision icon area on a single entry of Symptom Note the previously selected values for the items are displayed on the screen Empty value...

Page 9: ...data viewing screen is displayed upon tapping the red rectangular section No Description 1 ECG recording start time and end time 2 Move to ECG recording data viewing screen icon 3 Go Back icon 4 Move...

Page 10: ...could result 11 Cautions should be communicated to the patient 12 All serious accidents related to the medical devices should be reported to the health authorities of the Member State where the manufa...

Page 11: ...th waste disposal regulations may result in environmental pollution NOTE The data stored within the device must be taken care of before disposal Conditions for Use 1 Temperature range 10 45 2 Relative...

Page 12: ...Immunity Immunity Test IEC 60601 Test and Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 8 kV Contact 2 4 8 15 kV Air Floors should be wood concrete or...