Intended Purpose

  Device is intended to measure, analyze, and report electrocardiogram (ECG) information for long-term monitoring (up to 14 days) by attaching to the skin surface.

  It is used for patients with heart disease such as arrhythmia or patients with suspected heart disease. Reported ECG metrics include single lead analysis in beats, 

  heart rate measurement and rhythm analysis. Used by patients as prescribed by physician or medical personnel.

Target Treatment Group

- Patients with heart disease such as arrhythmia.
- Patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, presyncope, 

   syncope, fatigue, or anxiety.

Target User

  The patient (AT-Note & Device) 

Clinical Benefit 

  It measures patient electrocardiogram (ECG) information for a long-term monitoring, analyzes and reports measured data, and helps physician or medical

  personnel prescribe it.

  Disposal of Device and battery must comply with local waste disposal regulations. Non-compliance with waste disposal regulations may result in environmental 

  pollution. 
  NOTE : The data stored within the device must be taken care of before disposal.

Conditions for Use

  1) Temperature range : 10℃ ~ 45℃
  2) Relative humidity range : 10% ~ 95%, non-condensing 
  3) Atmospheric range : 700hPa ~ 1060hPa

Conditions for Storage

  1) Temperature range : -20℃ ~ 55℃
  2) Relative humidity range : 0% ~ 95%, non-condensing 
  3) Atmospheric range : 700hPa ~ 1060hPa
  4) Keep dry and keep away from sunlight. 

20

Clinical information

Product Disposal

Conditions for use and Storage

Appendix

04

21

Item

Description

ECG

Type of Applied Part 

BF type

Sampling Rate

250 sample / sec

Input Offset Dynamic Range

± 300mV

Channel

1 channel

ADC Resolution

10 bits

Input Impedance

>10㏁

Frequency Response

0.67Hz to 40Hz

Electrode

AC impedance

Less than 3KΩ (10Hz)

RF

RF communication

2.4GHz BLE 4.2

Effective Radiated Power

< 1mW

RF Frequency Band of TX

2.4GHz

Bandwidth of the Receiver

2400 ~ 2480MHz

Power

Requirement

Power Supply

DC 3V, Coin Battery (CR2032)

Physical

Characteristics

IP Classification

IP 57 

Total Size (L x W x H : mm)

84 x 48.8 x 8.5

Main Body Size (L x W x H : mm)

31 x 39 x 7.8

Sub Patch Size (L x W x H : mm)

104 x 68.8 x 0.05

Weight (g)

Main Body : Below 12g

Sub Patch : Below 1g

Specifications

Appendix

04

Summary of Contents for AT-Patch

Page 1: ...AT Patch User Manual P94A000001A DUG 02 EU Rev 0 02 2022 This product is a disposable medical device Reuse is prohibited...

Page 2: ...attach the device Additional Information 4 7 Caution Clinical information Specifications Electrical Safety and Compatibility Symbols 18 20 21 22 23 3 Introduction 01 Product Components Sub Patch 1EA M...

Page 3: ...least 2 minutes Cleaning the area where the device will be attached with alcohol How to attach the device How to use the device 02 How to use the device 02 5 2 3 4 Turning on the power Press the power...

Page 4: ...at once Reattachment can not be possible Attaching the Sub Patch 7 After attaching the Sub Patch detach the protective tape from the base film slowly Removing protective tape 8 Checking the attachment...

Page 5: ...ps then start using the App Supports Android 5 0 or later iOS 11 0 or later How to install the App 8 Screen with the App AT Note Icon Initial Run Screen How to use the App 03 How to use the App 03 No...

Page 6: ...e Symptom Note entry screen 6 A section displaying the ECG and 3 axis data output through graphs in real time LIVE Display Screen No Description No Description No Description 1 Stop generating real ti...

Page 7: ...the App receives the event information from the device and a dialog box is displayed to ask whether to record the symptoms immediately in a pop up window If the Later is tapped the Symptom Note entry...

Page 8: ...gistered at the App AT Note Upon tapping each revision icon or revision icon area on a single entry of Symptom Note the previously selected values for the items are displayed on the screen Empty value...

Page 9: ...data viewing screen is displayed upon tapping the red rectangular section No Description 1 ECG recording start time and end time 2 Move to ECG recording data viewing screen icon 3 Go Back icon 4 Move...

Page 10: ...could result 11 Cautions should be communicated to the patient 12 All serious accidents related to the medical devices should be reported to the health authorities of the Member State where the manufa...

Page 11: ...th waste disposal regulations may result in environmental pollution NOTE The data stored within the device must be taken care of before disposal Conditions for Use 1 Temperature range 10 45 2 Relative...

Page 12: ...Immunity Immunity Test IEC 60601 Test and Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 8 kV Contact 2 4 8 15 kV Air Floors should be wood concrete or...

Page 13: ...s Co Ltd All Rights Reserved ATsens Co Ltd Point Town 803 11 Gumi ro Bundang gu Seongnam si Gyeonggi do 13637 Republic of Korea Tel 82 70 5220 0220 CMC Medical Devices Drug S L C Horacio Lengo n18 C P...

Reviews: