ATORN DFT
Calibration
Display description - calibration step Zero
1 Measuring application (example)
2 Name of calibration (example)
3 Schematic illustration of the current
calibration step
4 Current calibration step
5 Progress display of the calibration
steps (example of 3 calibration steps,
calibration step 1 current)
6 Currently measured reading
(example)
7 Mean value of the existing number of
measured readings (example)
8 Delete last measured reading
9 Battery indicator
10 Number of measurements (example)
11 To next calibration step/exit calibration
7.3.5
Calibration step Foil
Measurement on the calibration foil lying directly on the uncoated reference
part.
Required material
•
Calibration foils with the desired thicknesses or from the scope of supply.
The circle on the foil marks the specified measurement area.
•
Reference part from customers own production, without the coating to be
measured
NF/FE
Calibration 003
Zero
X
OK
1
2
3
4
5
6
7
8
9
10
11