12
Fig. 8.
CPC-Coupling
Connection of the compressed-air hose (optional)
Connection of the function hose for compressed air at the
side of the column.
•
Compressed-air hose is connected by means of plug
connections; make sure that the plug attachment is
engaged. The hose is disconnected by pressing the
release slide (
, fi g. 8)
at the plug connector and by
pulling out the plug attachment.
Push the handles into the handle support as indicated
by the symbols.
The photoelectric barriers will operate the wrong
pump if the handles are transposed!
3.6
Starting-up the ATMOS
®
C 31
(contin.)
Connecting the hose for ear rinsing (optional)
The hose for the warm water system is screwed on the
CPC coupling.
Different socket nipples; misplacing of the hoses is
thus impossible !
Connecting the hose for ear rinsing (optional)
Connect the aqua stop valve with the water tap (local
connection). Open the water supply. When the main switch
is set to ON and a water consuming device (basin rinsing
or hose rinsing system) is used the water supply is opened
by automatic opening of the magnetic valve in the aqua
stop system.
Thus, there is no pressure inside the water supply
hose if no water consuming device is activated.
Fig. 9.
Connecting the hose for ear rinsing
Connection for compressed-air hose
Connection for ear rinsing hose
3.0 Installation and starting-up
Fig. 7a.
Connecting LED
Light package LED REF 506.7550.0 (Fig. 7a)
•
Connecting for the ATMOS LS 21 LED light
source and ATMOS HL 21 LED headlight.
Switch LED I and LED II
Holder for ATMOS HL 21 LED headlight
Holder for ATMOS LS 21 LED light source
Connecting socket LED II
Connecting socket LED I