GE
3.0 Setting up and starting
3.3
Explanation of the controls
Control light 1: Ready for use (green)
Control light 2: Fault (red)
Control light 3: Disinfectant sensor (yellow)
Alarm RESET button
Switch On/Checking routine: The green control signal
(Fig. 6) jitters:
During the system is switched on the ATMOS
®
Aqua clean
automatically performs a checking routine. While this
procedure the control signal (Fig. 6) is blinking fast.
If an error occurs the unit is not activated and the error is
shown at the controls (see below). After successful che-
cking routine, control signal 1 illuminates constant and the
unit is ready for operation.
The green control signal and the yellow signal
(Fig. 7) are illuminated and an alarm tone
sounds.
Cause:
Disinfectant sensor indicates, that the processing
water is not being sterilised!
Action:
Change the chemical bottle (see chapter 4.2.1)!
If the chemical bottle cannot be changed immediately,
the RESET button must be pressed to deactivate the
continuous alarm tone. The unit remains functional, and
does not interrupt the work. However, a short signal tone
still sounds periodically. If this error was only produced
by an air bubble, it will expire within the next trial by itself.
However, no decontamination takes place until the empty
chemical bottle is replaced with a full one (see chapter
4.2.1).
The green control signal is illuminated and the red
signal jitters and an alarm tone sounds (Fig. 8).
Cause:
Level monitor in the mixing container is soiled
or there is no water supply!
Action:
First of all, check if the main water tap resp. water
feed line is opened! Clean or replace the mixing container
probe (see chapter 4.2).
Consequence:
All valves and the pump are switched off.
The alarm tone can be switched off by the RESET button.
After solving the problem the ATMOS
®
Aqua clean can
only be activated again by switching the unit off and on.
The green control signal and the red signal
are illuminated and an alarm tone sounds (Fig. 9).
Cause:
Defective double membrane pump or overflow!
Action:
Clean the overflow sensor. Check the double
membrane pump and the electronic control system (see
chapter 4.2).
Consequence:
All valves and the pump are switched off.
The alarm tone can be switched off by the RESET button.
After solving the problem the ATMOS
®
Aqua clean can
only
be activated again by switching the unit off and on.