13
VORLA
GE
4.0 Cleaning and disinfection
4.2
Care and cleaning
Automatic emptying:
For emptying the unit during operation
the RESET button
4
has to be held down for approx. 8 sec.
(control signal
1
jitters) until a second acoustic signal is
activated. Then the green control signal 1 jitters slowly. Supply
container and pressure container are empty if no more water
comes out of the unit. Now, the ATMOS
®
Aqua clean can
either be switched off or the normal operation can be activated
by holding down the RESET button for another 8 sec.
Intensive decontamination:
To increase the concentration of
the chemicals as required, the intensive decontamination can
be activated by holding down the RESET button
4
(control
signal
1
jitters) for approx. 4 sec. until the first acoustic signal
is activated. While this procedure is taking place control signal
1
jitters and signal
3
is illuminated. Afterwards the handle
must be oprated. With the next chemical dosing operation
will be delivered 3 doses of chemical, followed by 2 doses
each in the next 4 dosing operations. No chemical will be
added to the completing dosing in order that the normal
concentration is ensured with finalisation of the intensive
decontamination. Control signal 3 expires after the intensive
decontamination and a short acoustic signal is activated. The
ATMOS
®
Aqua clean returns into normal operation. Intensive
decontamination is always recommended prior to longer idle
times like weekend or holiday and after treatment of patients
with suspicion of HIV, HBV or HCV.
Remove water level probe from the mixing container and
clean or replace.
●
Switch off main switch.
●
Carefully pull probe up and out.
●
Clean or replace probe. Check the electronic control
system.
●
Replace the probe in the mixing container
1
.
●
Switch on main switch.
Check overflow sensor at the mixing container.
●
Switch off main switch.
●
Clean/dry overflow sensor
2
. Check electrical
connections of the electronic control system
(see page 22).
●
Check double membrane pump and the electronic
control system.
●
Switch on main switch.
If this error exists further on, the double membrane is
probably broken, whereby the double membrane pump
(module 5) has to be replaced!
Surface disinfection
Please use ATMOS
®
Green & Clean SK for the daily
surface disinfection .
Replacing the filter element
Replace the filter element every three months or when
it is contaminated or discoloured. Always wear new, dry
disposable gloves when replacing the filter. Pay attention
that the filter element does not come into contact with wet or
contaminated objects. Rinse the filter housing thoroughly and
disinfect it with alcohol free, ph-neutral disinfectant.
The filter element is a single-use product. The filter elements
must be stored in a dust-proof and dry place.
1
2
3
4
1
2
Fig. 19
Fig. 20