Q46C2 2Electode Conductivity System
Part 1 - Introduction
8
O&M Manual
Rev-H (6/19)
Max. Sensor-Instrument
60 ft. (18.2 meters)
Distance
Sensor Types
Model Q25C2 - 4 wire input or equal
Accuracy Note
:
The Q46C2 is designed to allow measurements over a broad conductivity range using sensors with
either 0.05K or 0.5K cell constants. However, extending the measurements made with various cell
constants can result in some non-linearity at the high end of operating ranges. This can reduce
accuracy from 0.1% to 1-2% in some cases. When high accuracy is critical, the following provides
a list of the preferred cell constant to be used for various operating ranges in order to maintain the
highest accuracy.
0.05K Sensor
– Use for all resistivity ranges (20 M
or 200 K
)
0.05K Sensor
– Use for 0-2.000 or 0-20.00 uS ranges.
0.5K Sensor
– Use for 0-20.00, 0-200.0, 0-2000 uS ranges.