ATI
Q46C2 2Electode Conductivity System
Part 4
– Electrical Installation
20
O&M Manual
Rev-H (6/19)
4.6
Direct Sensor Connection
Sensor connections are made in accordance with Figure 14. The sensor cable can be routed into
the enclosure through one of cord-grips supplied with the unit. Routing sensor wiring through
conduit is only recommended if a junction box is to be used. Some loose cable is needed near the
installation point so that the sensor can be inserted and removed easily from the flowcell.
Cord-grips used for sealing the cable should be snugly tightened after electrical connections
have been made to prevent moisture incursion. When stripping cables, leave adequate length for
connections in the transmitter enclosure as shown below. The standard 15 ft. sensor cable
normally supplied with the system is already stripped and ready for wiring. This cable can be cut
to a shorter length if desired to remove extra cable in a given installation. Do not cut the cable so
short as to make installation and removal of the sensor difficult.
.
Figure 13 - Sensor Cable Preparation