ATI Technologies Q45H/62 Manual Download Page 6

Analytical Technology Unit 1 & 2 Gatehead business PK, OL3 5DE 
Phone +44 145 783 2800 
Fax +44 145 783 9500 
[email protected]

 

 

Q45H Training Handout 4.05 

Page 5 of 11

 

Short Guide for Calibration of pH corrected Q45H62 chlorine monitor  

This guide is intended for use for routine calibration only. If problems during calibration occur please consult the 
full O&M manual. Main thing to remember is to calibrate the pH using 2 standards prior to calibrating the chlorine 
(only when the correction is turned ON).  

 

 

General 

To calibrate the Q45H62 Residual Chlorine Monitor with pH correction turned ON you will need: 

 

1.

 

pH buffers. Ideally pH 4 and pH 7. On applications with pH 7+ use buffers 7 and 10. 

2.

 

Method of determining the level of residual chlorine, e.g. Hach colorimeter or Palin test. 

3.

 

Chlorinated water 0. for calibration (this minimum is required due to potential error of dpd test. 

 

pH Calibration (Must do first) 

 

Keys to Press 

Screen Display 

User Action Required 

MENU/ESC until 

CAL MENU 

 

ENTER 

Cal Cl2 

 

 

Cal Temp, Set Range, 
Cal pH type. 2 Pnt,  Cal 
pH buf1 

If Cal pH type 1 Pnt is displayed.  
Press ENTER and use

  to select 2 

Pnt 

Place sensor in buffer 1  

 

 

Wait 1 – 2 minutes (stir probe) 

 

 

ENTER 

Cal pH buf XX.X pH 

Right hand digit flashes 

Adjust pH value using left arrow to 
select digit and to 

 adjust 

 

 

ENTER 

Accepted 

 

Rinse sensor then place in buffer 2 
(low) 

 

Wait 1 minute 

10 

 

Cal pH buf2 

 

11 

ENTER 

Cal pH buf XX.x pH 

Right hand digit flashes 

12 

Adjust pH value using left arrow to 
select digit and to 

 adjust 

 

 

14 

ENTER 

Accepted 

 

15 

To return to MEASURE Menu use 
MENU/ESC, to go to CAL Cl2 press 

 

 

 

Normal (“1 Ref”) Cl

2

 calibration  

 

Keys to Press 

Screen Display 

User Action Required 

a) 

 or MENU/ESC 

CAL MENU 

 

ENTER 

Cal Cl2 

 

ENTER 

Set cal mode, 1 ref 

If the display says 2 Zero or 3 cmp 
press ENTER and use 

  to adjust to 1 

Summary of Contents for Q45H/62

Page 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Page 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Page 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Page 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Page 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Page 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Page 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Page 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Page 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Page 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Page 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...