ATI Technologies Q45H/62 Manual Download Page 10

Analytical Technology Unit 1 & 2 Gatehead business PK, OL3 5DE 
Phone +44 145 783 2800 
Fax +44 145 783 9500 
[email protected]

 

 

Q45H Training Handout 4.05 

Page 9 of 11

 

 

Quick Reference Guide (PRINT OFF AND DISPLAY) 

 
Observation Checklist 
 
1.

 

Flow through the cell going over weir. 

2.

 

Check the pH reading is the same and correct. 

3.

 

Try to calibrate span (REF 1) 

 

Scroll to the CAL menu section and press ENTER. Cal Cl2 will be displayed. 

 

Press the ENTER key. The screen will display a flashing 1-Ref for span calibration. 

 

Press the ENTER key. Waiting will be displayed. 

 

Take  a  hand  held  chlorine  measurement  and  the  last  ppm  measurement  will  be  display.  Using 
the arrow keys to adjust to the new reading. 

 

Press the ENTER key. If accepted the display the message PASS. 

 

 

The sensor offset value is display in % from the last calibration

If  the  calibration  fails,  a  message  indicating  the  cause  of  the  failure  and  the  FAIL  icon  will  be 
turned on.   

 

4.

 

If this is accepted the unit just needed to be calibrated.  

5.

 

If this is not accepted.  

6.

 

Note the sensitivity of the sensor. 

7.

 

The range of acceptable values for slope is 20-500% (if the sensitivity<12%). It will be necessary to 
rebuild sensor.  

8.

 

Inspect the sensor end is it white or dirty. 

 
Rebuilding Sensor  
1.

 

Unscrew the electrolyte chamber from the assembled sensor and also remove the fill screw from the 
side of the sensor body. 

2.

 

Remove  the  front  nut  from  the  bottom  of  the  chamber  and  discard  the  protective  membrane.    O-
rings  are  contained  in  grooves  on  both  the  bottom  and  top  of  the  chamber.    Be  sure  that  these  o-
rings remain in place. 

3.

 

From the package of membranes supplied with the sensor, place a new membrane into the front nut.  
The  membrane  is  white  in  colour  and  is  separated  from  other  membranes  by  a  light  blue  paper 
spacer. 

4.

 

Screw the front nut on to the chamber until you feel the o-ring compress.  Hand tight compression is 
all that is needed.  Do not use tools to tighten.  The membrane should be flat across the bottom of 
the chamber without wrinkles. 

5.

 

Fill the chamber with electrolyte until the level reaches the bottom of the internal threads. 

6.

 

Slowly  screw  the  chamber  onto  the  sensor  body.    A  small  amount  of  electrolyte  will  run  out  of  the 
hole  from  which  the  fill  screw  was  removed.    Place a  paper  towel  around  the  sensor  to  absorb  the 
electrolyte overflow.  The electrolyte is harmless and will not irritate skin.  Tighten the chamber until 
the o-ring at the top of the chamber is compressed.  Once again, do not use tools to tighten. 

7.

 

Shake excess electrolyte from the fill hole on the side of the sensor and replace the fill screw. 

 

CAUTION: 

When handling the assembled sensor, do not set the sensor on its tip or damage to the membrane will result.  Severe 

impacts on the tip of the sensor from dropping or other misuse may cause permanent damage to the sensor. 

 

 

Summary of Contents for Q45H/62

Page 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Page 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Page 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Page 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Page 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Page 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Page 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Page 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Page 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Page 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Page 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...

Reviews: