ATI Technologies Q45H/62 Manual Download Page 5

Analytical Technology Unit 1 & 2 Gatehead business PK, OL3 5DE 
Phone +44 145 783 2800 
Fax +44 145 783 9500 
[email protected]

 

 

Q45H Training Handout 4.05 

Page 4 of 11

 

Chlorine Span Cal 

“1 Ref” 

Span calibration of the system must be done against known value of chlorine in the sample ‘at the point in which you 
are calibrating the unit and no longer than a couple of minutes to prevent an alteration in the process that could lead 
to an incorrect cal’.  A sample should be collected from the inlet line feeding the flow cell and quickly analyzed for 
free chlorine. The chlorine monitor should be calibrated while operating on a chlorinated sample stream in the flow 
cell assembly.  
 

IS YOUR SAMPLE WATER > 8 PH AND VARIABLE OF +/- 0.2 (i.e. 8.3 today and 8.5 tomorrow) PLEASE 

CONTACT ATI FOR ASSISTANCE WITH CONFIGURATION OF THE INTERNAL CURVE CORRECTION! 

 

ATi uses a DPD colorimetric method to measure the free chlorine. Calibrate system as follows: 

 

1.

 

Once the sensor has been in the flow cell long enough for polarization (8 hours on a new sensor, 15-20 minutes 
on an already ‘polarized’ sensor) you can begin the calibration. If the sensor is on-line the user may want to set 
the output HOLD feature prior to calibration to lock out any output fluctuations. 

 
2.

 

Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.  

Cal Cl

2

 will then be 

displayed. 

 
3.

 

Press the ENTER key.  The screen will display a flashing 

1-Ref

 for span calibration or a 

2-Zer

 for zero 

calibration. Using the UP arrow key, set for a 

1-Ref

 span calibration and press ENTER. AT THIS POINT CARRY 

OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET 
THE CORRECT AVERAGE! 

 
4.

 

The system now begins acquiring data for the calibration value.  As data is gathered, the units for ppm and 
temperature may flash.  Flashing units indicate that this parameter is unstable.  The calibration data point 
acquisition will stop only when the data remains stable for a pre-determined amount of time. This can be 
overridden by pressing ENTER twice.  If the data remains unstable for 10 minutes, the calibration will fail and the 
message 

Cal

 

Unstable

 will be displayed. 

 
5.

 

The screen will display the last measured ppm value and a message will be displayed prompting the user for the 
DPD value.  The user must then modify the screen value with the arrow keys and press ENTER.  The system then 
performs the proper checks.  

 
6.

 

If accepted, the screen will display the message 

PASS

 with the new sensor slope reading, and then it will return 

to the main measurement display.  If the calibration fails, a message indicating the cause of the failure will be 
displayed and the FAIL icon will be turned on.  The range of acceptable values for sensor slope is 20% to 600%. 
It may be necessary to rebuild the sensor as described on page 2, Chlorine Sensor Assembly.  Should the slope 
value remain out of range and result in calibration failures, review the Service Section of this manual, then 
contact the service dept. at ATI for further assistance. 

 

The sensor offset value in % from the last span calibration is displayed on the lower line of the Default Menus for 
information purposes.  Once the % goes to <25% THE NEXT TIME YOU CAL CHANGE MEMBRANE & 
ELECTROLYTE AS THE SLOPE IS AT ITS MINIMUM BEFORE FAILING.

 

 

Summary of Contents for Q45H/62

Page 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Page 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Page 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Page 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Page 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Page 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Page 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Page 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Page 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Page 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Page 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...

Reviews: