ATI Technologies Q45H/62 Manual Download Page 1

 
 
 

Model Q45H/62 

 

Free Chlorine 
Measurement 

System 

 

Training Handout

 

 
 
 
 
 
 
 
 
 
 
 
 

European Office 

 
Analytical Technology 
Unit 1&2, Gatehead Business Park 
Delph New Road, Delph 
Saddleworth 
OL3 5DE 
Tel     01457 873 318 
Fax   01457 874 468 
Email  [email protected]

 

Summary of Contents for Q45H/62

Page 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Page 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Page 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Page 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Page 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Page 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Page 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Page 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Page 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Page 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Page 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...

Reviews: