7-2-2
Viscosity Input
[Scale Selection Screen]
Slide
to select a scale
number, then push
to
confirm the selection.
20.0
°
C
Scale Selection
01 05 09
02 06 10
03 07
04 08
20.0
°
C
Scale Setting 01
X1: 5449.0mPas
X2: 4451.0mPas
X3: 4157.0 mPas
Next>
[Scale Setting (X) Screen]
Input the three
measurement values from
the VISCO.
Slide
to select an
item, then push
to
continue to the input screen.
20.0
°
C
4157.0
1 2 3 4 5 6 7 8
9 0 . ← →
Back
Enter Quit
20.0
°
C
Scale Setting 01
Y1: 5810.0mPas
Y2: 4763.0mPas
Y3: 4400.0 mPas
<Back Enter
[Scale Setting (Y) Screen]
Input the three
measurement values from
the other viscometer.
Slide
to select an
item, then push
to
continue to the input screen.
[Main Menu Screen]
Save
[Scale Setting (X) Screen]
20.0
°
C
4400.0
1 2 3 4 5 6 7 8
9 0 . ← →
Back
Enter Quit
[Input Screen]
[Input Screen]
m
Turning the user scale function ON/OFF
and selecting a scale number is done from
the measurement screen (
C
p.30).
m
If attempting to input the same viscosity for
X1, X2, X3, a beeping sound will indicate that
this setting is invalid.
If attempting to input the same viscosity for
Y1, Y2, Y3, a beeping sound will indicate that
this setting is invalid.
N
Select "Quit" to cancel
settings and return to the
Scale Setting screen.
N
Values up to 1 decimal
place can be input.
←→: use the arrows to
move the cursor.
Back: select "Back" to
delete the value to the
left of the cursor.
Slide
to select a number
value, then push
to input it.
20.0
°
C
4157.0
1 2 3 4 5 6 7 8
9 0 . ← →
Back
Enter Quit
Input
Field
About the input screen