![ATAGO VISCO Quick Manual Download Page 23](http://html.mh-extra.com/html/atago/visco/visco_quick-manual_2995845023.webp)
18
3-2-7
Measurement
Removing the spindle:
While pressing on the shaft
holder, pull the spindle
straight out.
If changing the spindle:
1.
Remove the spindle and temperature sensor.
2.
Attach the spindle you wish to use and the
temperature sensor. Be sure to wash the temperature
sensor before re-attaching (
C
p.14).
m
Make sure no sample residue is left on the
shaft holder or temperature sensor
attachment point.
m
If sample residue is present on the spindle and temperature sensor, wash them
thoroughly.
m
After changing the spindle, check that the surface of the sample is level with the
spindle's reference line.
20.0
°
C
--
A1L 100.0rpm
0.0
mPas
Max:
1,657mPas
Start
Torque % bar
[Measurement Screen]
Slide
to select
"Start," then push
to confirm the selection
and begin measurement.
Slide
to select
"Quit," then push
to confirm the selection
and terminate
measurement.
[Screen During Measurement]
20.0
°
C
--
A1L 200.0rpm
558.2
mPas
67.4%
Quit
Viscosity
Speed
N
Settings can be changed
even during measurement.
Torque %
N
The recommended
range is 10.0 to 100.0% (and 5.0
to 100.0% when using Ultra Low
Adapter (ULA)-Sample
Adapter for Low Viscosity
Sample (sold separately)).
If the value falls outside of the
above range, a "beep" will be
emitted.(
C
p.39)
Shaft holder
Summary of Contents for VISCO