14-4
Conveyors
Checking Part Sensors
To check the part sensors:
1.
In the FmNT Main Window, click on
Tools
.
2.
Click on
I/O Test
.
3.
Click on
Conveyor
.
>
This opens the Test Window, which displays a list of all inputs and outputs.
4.
Turn the Interlock ON (
I
).
5.
Press
Pause
on the Control Panel.
WARNING!
Wait for all Dispensing Head and conveyor motion to stop before
opening the doors.
6.
Open the doors.
7.
Manually move a part into the Pre-dispense or Dispense Zone until it is directly over the sensor.
>
For left-to-right conveyors, watch the
Left Sense I/O
in the Test Window.
>
For right-to-left conveyors, watch the
Right Sense I/O
in the Test Window.
>
If either the
Left
or
Right Sense I/O
does turn ON, the sensor is functioning.
>
If both the
Left
and
Right Sense I/O
fail to turn ON, the sensor is not functioning.
Call an Asymtek service technician.