P72
Electromagnetic immunity
All
LU700 series products are in compliance with the regulation of immunity test,
and the detail and declaration as below:
M
anufacturer’s declaration-electromagnetic immunity
The LU700 Series is intended for use in the electromagnetic environment (for professional healthcare) specified below.
The customer or the user of the LU700 Series should assure that it is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
(for professional healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
:
0,15 MHz
– 80 MHz
6 Vrms
:
in ISM bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz
– 2,7 GHz
80 % AM at 1 kHz
3 Vrms
:
0,15 MHz
– 80 MHz
6 Vrms
:
in ISM bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
e)
3 V/m
80 MHz
– 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the
LU700 Series
including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = 1,2
√P
d = 1,2
√P
80MHz to 800 MHz
d = 2,3
√P
800MHz to 2,7 GHz
Where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter
manufacturer
and
d
is
the
recommended separation distance in meters
(m).
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.