Astrol Electronic BAT-FLEX Manual Download Page 7

 

 

Astrol Electronic AG reserves the right to change specifications without notice

 

 

 

Bat FLEX Manual

 

7/19

 

AD-10069-007 Rev. 03 

2.5.3.

 

BatFlex Reference Card 

2.5.3.1.

 

Specification 

Channels

 

Details 

Value 

Unit 

Number of channels 

10 

Pcs. 

2.5.3.2.

 

Warnings 

 

The  output  cables  must  not  touch  the  case  in  any  way.  Connecting  the  reference  output  to  the  case  will  lead  to  wrong 
calibration of the device. 

 

2.5.4.

 

BatFlex Small Dual Card 

For informations about the output connector please read Appendix B: BAT FLEX Small Connection Designation 

2.5.4.1.

 

Specification 

Channels

 

Details 

Value 

Unit 

Number of channels 

Pcs. 

 

Current measurement

 

Details 

Value 

Unit 

Resolution 

± 188 uA, 

100 

pA 

 

± 750 uA 

400 

pA 

 

± 3 mA 

1.5 

nA 

 

± 24 mA 

12 

nA 

Accuracy 

full scale 

±0.05 

Long term drift 

 

~30 

ppm / day 

 

Current output

 

Details 

Value 

Unit 

Output Range 

Resolution: 50 nA 

±24 

mA 

Voltage Range 

Current = 24mA 

± 10 

 

Voltage Measurement 

Details 

Value 

Unit 

Range 

 

± 12.5 

Resolution 

 

μV 

Accuracy 

full scale 

± 0.01 

Long term drift 

 

~10 

ppm / day 

 

2.5.4.2.

 

Warnings 

 

Do not short circuit different guard pins. This can cause permanent damage to the card. 

 

The  reference  output  cable  must  not  touch  the  case  in  any  way.  Connecting  the  reference  output  to  the  case  will  lead  to 
wrong calibration of the device. 

 

 

Summary of Contents for BAT-FLEX

Page 1: ...e specifications without notice Bat FLEX Manual 1 19 AD 10069 007 Rev 03 BAT FLEX Manual Rev Remarks changes created checked released 01 Initial ported RT 01 03 13 AST 01 03 13 AST 01 03 13 03 Correct...

Page 2: ...ard 7 2 5 3 BatFlex Small Dual Card 7 3 CCCC Setup 8 3 1 Preparation 8 3 2 Communication Setup 8 3 3 CCCC Overview 9 4 Set up the Measurement Program 10 4 1 Introduction 10 4 2 Measurement Flow 11 4 3...

Page 3: ...cription how to set the device IP can be found in the BatFlex FPGA Card Menu Guide Ethernet User PC 1 2 Direct Connection Please read Appendix A How to set up a static IP for further instructions User...

Page 4: ...ected or unavoidable maintenance of the network could cause measurement data loss 2 3 Card Types The BatFlex Device is designed as amodular System where measurement cards can be replaced whenever the...

Page 5: ...by turning the wheel to the left or to the right Selections are confirmed by pressing the wheel The SD card slot can be used to upgrade the firmware of the device USB and JTAG are for service and deb...

Page 6: ...ice FPGA IMG from SD A new FPGA image will be transferred from the SD card to the device This is a major update which means that the entire software is renewed Firmware from SDCard The FPGA boots up s...

Page 7: ...gnation 2 5 4 1 Specification Channels Details Value Unit Number of channels 8 Pcs Current measurement Details Value Unit Resolution 188 uA 100 pA 750 uA 400 pA 3 mA 1 5 nA 24 mA 12 nA Accuracy full s...

Page 8: ...needs to be restarted On startup the application searches for a viable device on the Network If the device is not yet configured the first action after the main window shows up is to get into the comm...

Page 9: ...esn t switch automatically to constant current mode The range can be chosen by clicking on I and selecting the appropriate range in the dropdown menu Output Modes Constant voltage By clicking on Uout...

Page 10: ...is created by the CCCC Software during a measurement It contains the measurement data recorded by the device Log File log This File contains all the events that took place during the measurement For e...

Page 11: ...en move from position to position during the execution Each step has its unique ID and can be configured with individual parameters It s also possible to jump between the steps by setting conditions t...

Page 12: ...ostatic Step This step implements a galvanostat or current source Its main purpose is to keep the current through the cell constant disregarding changes in the load itself This step offers a condition...

Page 13: ...channel has to be selected After that the measurement window can be opened The measurement window contains control elements to start stop and pause the data recording as well as a table with the curre...

Page 14: ...irst step is a counter step which sets the cycle counter to one The following two steps are galvanostatic steps which charge discharge with a constant current until the defined end voltage is reached...

Page 15: ...Tool offers two main methods to display the data as a CV diagram Current Voltage or as a time dependent diagram There are many customizable parameters to get the best view on the desired data 7 Expor...

Page 16: ...twork card with up to date network drivers Open the Network and Sharing Center This can be found in the Control Panel under Network and Internet A Control Panel can be opened by pressing Windows R the...

Page 17: ...ddress This will activate the input fields One should use the data obtained in the first step to complete the Subnet mask Gateway and the DNS server addresses The computer will change the IP address a...

Page 18: ...e 1 Guard 2 Never connect different guard types the device could be damaged 2 Aux 2 3 Guard 2 4 Guard 2 5 Aux 2 6 Guard 2 7 NC Not connected 8 Guard 1 9 Aux 1 10 Guard 1 11 Guard 1 12 Out 1 13 Guard 1...

Page 19: ...l with serial number 11 2 Inside Type label with serial number on the left side 12 Order code Type Order Code BatFlex Base Device with Reference Card AA 10320 001 Bat Flex Small Dual 8 Channels AA 103...

Reviews: