background image

6

BACK PANEL CONFIGURATIONS

The modular design allows for a variety of configurations. In addition to master 
or slave configurations, the scanners can also be set-up with the following 
configurations: 8 or 16 High Voltage testing channels, 8 High Voltage and Ground 
Bond testing channels, and 8 or 16 Ground Bond testing channels.

Model SC6540 HNM

8 Channel High Voltage Scanner

Model SC6540 HGM

8 Channel High Current Scanner  
8 Channel High Voltage Scanner

Model SC6540 GGM

16 Channel High Current Scanner

*

Also available in slave configurations

The different configurations (shown right)
are indicated by the following alpha
designators.

M

  =  Master Scanner

H

  =  8 High Voltage Channels

HH

 =  16 High Voltage Channels

G

  =  8 Ground Bond Channels

GG

 =  16 Ground Bond Channels

N

  =  Empty Module

S

  = Slave

Model SC6540 HHM

16 Channel High Voltage Scanner

Model SC6540 GNM

8 Channel High Current Scanner

Summary of Contents for SC6540

Page 1: ...always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING AND USING A ...

Page 2: ...le is a High Voltage Continuity module If the green LED illuminates it indicates that the installed module is a Ground Bond module MODULE B CHANNEL STATUS INDICATORS These LED s indicate the status of each individual channel on Module B If the red LED illuminates it indicates a High Voltage Continuity Current Ground Bond channel If the green LED illuminates it indicates a Return channel MODULE A C...

Page 3: ...Module SCANNER BUS OUTPUT Interconnect port for the control cable to another SC6540 in a multiple SC6540 system ADDRESS SWITCHES 8 pin DIP switch used to address the modules in a SC6540 slave or used to configure the address of a SC6540 master CURRENT INPUT JACK Connector used to attach the high current input lead or Continuity Current input lead from the host instrument RETURN INPUT Connector for...

Page 4: ...he fuse compartment will be exposed Replace the fuse with one of the proper rating INPUT POWER RECEPTACLE Standard IEC 320 connector for a standard NEMA style line power mains cord POWER SWITCH Rocker style power switch with international ON and OFF 0 markings INPUT VOLTAGE SWITCH Line voltage selection is set by the position of the switch In the left position it is set for 110 120 volt operation ...

Page 5: ...andard GPIB interface or Ethernet This model receives control information from a PC and can also deliver instructions to up to four additional slave Scanners A master Scanner can be distinguished by its power module located on the upper left side of the rear panel SLAVE A slave Scanner only receives data The data that the slave receives can come from a master Scanner remote control or directly fro...

Page 6: ...40 HNM 8 Channel High Voltage Scanner Model SC6540 HGM 8 Channel High Current Scanner 8 Channel High Voltage Scanner Model SC6540 GGM 16 Channel High Current Scanner Also available in slave configurations The different configurations shown right are indicated by the following alpha designators M Master Scanner H 8 High Voltage Channels HH 16 High Voltage Channels G 8 Ground Bond Channels GG 16 Gro...

Page 7: ...LECTOR 115V 250V 2AMAX 50Hz 60Hz CH1 H V CH2 CH3 CH4 CH5 CH6 CH7 CH8 CAUTION HIGH VOLTAGE 5 0 KY MAX CAUTION HIGH VOLTAGE 5 0 KY MAX CURRENT CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 115 1 2 3 4 5 6 7 8 ON HypotULTRA SC6540 Master To PC SENSE RETURN CURRENT SENSE SCANNER 2 SCANNER 1 T6 3A 250V GND 100 120 200 240V 700VA MAX 50Hz 60Hz SIGNAL INPUT CAL SIGNAL OUTPUT CAUTION HIGH VOLTAGE 5KVAC MAX 6KVDC MAX RE...

Page 8: ...ternal channels The Int Scanner parameter in the menu pertains to an internal 8 channel Scanner and the ExtScanner1 parameter pertains to an 8 channel high voltage SC6540 Scanner To set the Scanner channels use the back and forward arrows and set the channels to High H Low L or Open O Use the enter key to save the values and move on to the next test parameter H High High voltage output channel for...

Page 9: ...el is set as Return the green LED will light up If the High Voltage channel is set to Open no LED will light up Once the SC6540 is incorporated into a test system it will act as an extension of the HypotULTRA The outputs will only activate while a test is being performed and will deactivate when the test is not running When a failure is detected the test will stop the output will be deactivated an...

Page 10: ...he wire lengths going from the SC6540 High Current output and the High Current Return should be kept as short as possible to limit the effect of test lead resistance OMNIA II SENSE H V SENSE CURRENT RETURN SIGNAL INPUT SIGNAL OUTPUT SCANNER 1 SCANNER 2 RS232 T10A 250V GND CAT II 115V 230V 10A MAX 50Hz 60HZ CAUTION HIGH VOLTAGE 5KV MAX CAUTION HIGH VOLTAGE 5KV MAX GROUND TEST OUTPUT CH H V TEST OUT...

Page 11: ... CAUTION HIGH VOLTAGE 5 0 KY MAX CURRENT 1 2 3 4 5 6 7 8 ON SC6540 Slave OMNIA II SENSE H V SENSE CURRENT RETURN SIGNAL INPUT SIGNAL OUTPUT SCANNER 1 SCANNER 2 RS232 T10A 250V GND CAT II 115V 230V 10A MAX 50Hz 60HZ CAUTION HIGH VOLTAGE 5KV MAX CAUTION HIGH VOLTAGE 5KV MAX GROUND TEST OUTPUT CH H V TEST OUTPUT OUTPUT REMOTE INTERFACE INPUT G COM CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 COM CH1 CH2 CH3 CH4 C...

Page 12: ... for a High Voltage test or a current return for a Ground Bond or Continuity test O Open Channel is neither an output nor a return OPERATING THE SC6540 WITH OMNIA II 8204 Continued SETUP The SC6540 Scanner channels can be set as High H for High Voltage or Continuity testing output Low L for the Return connection or Open O for OFF Channel setup is done using the OMNIA II menu or accompanying automa...

Page 13: ...ll then begin to test from the first step in the program Slave Scanner Power Once the SC6540 slave is connected to the OMNIA II the power on LED will light as soon as the power switch of the OMNIA II is turned on Master Scanner Power The SC6540 master is powered on by flipping the switch on the rear panel of the unit to the ON position LED Indicators The two leftmost LEDs for Module A and Module B...

Page 14: ... included communications cable 38592 Connect the Probe HI output on the panel of the instrument to the Current input on the panel of the Scanner Connect the Probe LO output on the panel of the instrument to the Return input on the panel of the Scanner Finally plug the DUT into the universal adaptor box EXTERNAL MD INTERNAL MD MODULE 277VAC MAX 30A MAX L N L PROBE HI GND CASE CURRENT RETURN PROBE L...

Page 15: ... L low or O off Press the Left and Right Arrow keys to move the highlighted cursor to the corresponding Scanner channel Setting a channel to H will connect the Probe HI of the 620L to the corresponding Scanner channel Setting a channel to L will connect the Probe LO of the 620L to the corresponding Scanner channel The operator may set the Scanner on a test by test basis For example for test step 0...

Page 16: ...ditional information about these and other key features of the SC6540 please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 Associated Research arisafety com 5 2018 FOLLOW US ...

Reviews: