SAFETY
4
Power
Ground Bond Test Equipment must be connected to a good ground. Be certain that the
power wiring to the test bench is properly polarized and that the proper low resistance
bonding to ground is in place.
Power to the test station should be arranged so that it can be shut off by one prominently
marked switch located at the entrance to the test area. In the event of an emergency,
anyone can cut off the power before entering the test area to offer assistance.
Work Area
Perform the tests on a nonconducting table or workbench, if possible.
There should not be any metal in the work area between the operator and the location
where products being tested will be positioned.
Position the tester so the operator does not have to reach over the product under test to
activate or adjust the tester. If the product or component being tested is small, it may be
possible to construct guards or an enclosure, made of a non-conducting material such as
clear acrylic, such that the item being tested is within the guards or enclosure during the
test, and fit them with switches so that the tester will not operate unless the guards are in
place or the enclosure closed.
Keep the area clean and uncluttered. All test equipment and test leads not absolutely
necessary for the test should be removed from the test bench and put away. It should be
clear to both the operator and to any observers which product is being tested, and which
ones are waiting to be tested or have already been tested.
Do not perform Ground Bond Tests in a combustible atmosphere or in any area where
combustible materials are present.
TEST OPERATOR
Qualifications
The operator should understand the electrical fundamentals of voltage, current, and
resistance.
Safety Procedures
Operators should be thoroughly trained to follow these and all other applicable safety rules
and procedures before they begin a test. Defeating any safety system should be treated as
a serious offense and should result in severe penalties, such as removal from the Ground
Bond Testing job. Allowing unauthorized personnel in the area during a test should also
be dealt with as a serious offense.
Dress
Operators should not wear jewelry which could accidentally complete a circuit.
Medical Restrictions
This instrument should not be operated by personnel with heart ailments or devices such
as pacemakers.