66
Volt-LO
This message appears on the display if the DUT/IP voltage drops below the Volt-LO
Trip setting.
Amp-HI
This message appears on the display if the current exceeds the Amp-HI Trip setting
Amp-LO
This message appears on the display if the current drops below the Amp-LO Trip
setting.
Line-OC
This message appears on the display if the 40 Amp < 1 s, short circuit protection is
tripped.
Watt-HI
This message appears on the display if the DUT input wattage exceeds the Watt-HI
Trip setting.
Watt-LO
This message appears on the display if the DUT input wattage drops below the Watt-
LO Trip setting.
PF-HI
This message appears on the display if the Power Factor exceeds the PF-HI Trip
setting.
PF-LO
This message appears on the display if the Power Factor drops below the PF-LO Trip
setting.
Leak-HI
This message appears on the display if the enclosure to neutral leakage current
exceeds the Leak-HI Trip setting.
Leak-LO
This message appears on the display if the enclosure to neutral leakage current drops
below the Leak-LO Trip setting.
Neutral-V
This message appears on the display if there is voltage present on the Neutral of the
DUT input receptacle.
OTP