123
However, the presence of this response does not mean that the instrument (in
the case of these commands only) completed the command. These
commands require a restarting of the hardware that controls the Ethernet
Protocols. Because of this, the user must wait before the Ethernet Card will
respond to another command. See the table below for the approximate wait
times necessary after one of the commands in the table is sent. In addition, the
current socket connection between the user’s terminal and the Ethernet Card is
no longer valid, and the user will need to close their current connection and
establish a new one.
Ethernet Card Settings Command Wait Times
IP Mode
Command
Wait Time After Command is
Sent*
Manual
SIA, SGA, SSM
8 seconds
SIM 0
14 seconds
Auto
SDN
14 seconds
SIM 1
8 seconds
*Wait times are approximate and can vary based on the user’s network.
07 Data Storage Card (DISCONTINUED)
This option gives the user capabilities for serial communication, barcode scanning,
and saving test results onto a USB Flash memory drive.
The Data Storage Card has three input/output ports. These ports are shown in Figure
3.
Figure 3.
The USB-A port is used for the connection of the USB Flash memory drive.
Note:
Please use the USB Flash memory drive that was supplied with the unit. Proper
file transfer can only be guaranteed with the use of this drive.
The port labeled
“Barcode” is a PS/2-type connector that is used for the connection of a barcode
scanner. The 9-pin D-
type subminiature connector labeled “RS232/RS485” is for the
connection of the 620L to an RS232 or RS485 communication bus. To configure all of
the features of the Data Storage Card, use the Data Card Settings menu.