3
Routine Maintenance
24
STERRAD
®
Sterilization System Routine Maintenance Guide
1. The Y changes to N automatically when a PM is due. The level of
PM due is indicated on the print-out after a cycle has been
completed, and on the display in the idle state. You can reset the N to
Y by pressing
OPEN DOOR
; XXXX automatically changes to the
level 1 or 2 PM service interval in cycles. If no PM is due, then the
cycles until the next level 1 or 2 PM are displayed.
2. Press
OPEN DOOR
to store the data, the total machine cycles, and
print the information while exiting the Edit mode. The maximum
storage capacity 10 PM services. When the 11th PM service record is
entered, the system deletes the oldest record, re-sequences the
existing records and enters the new record in the 10th position. The
system prints the PM service history when it is updated:
3. Press CANCEL to transfer to the PM Level 1 Cycles Interval menu.
PM Level 1 Cycles Interval
The following is displayed:
PM 1 CYCLES = XXXX
OPEN(chng)-CLOSE(accept)
XXXX is the current Level 1 PM cycles service interval.
1. Press
CLOSE DOOR
to move the cursor to each digit in turn,
starting at the rightmost digit, moving left, and wrapping back to the
rightmost digit.
2. Press
OPEN DOOR
to increment the digit by 1 ranging from 0 to 9.
The Level 1 PM cycles can be programmed to a value between 500
and 2000 cycles, with a default value of 1500 cycles.
3. Press
CANCEL
to transfer to the PM Level 2 Cycles Interval menu.
PM Level 2 Cycles Interval
The following is displayed:
PM 2 CYCLES = XXXX
OPEN(chng)-CLOSE(accept)
XXXX is the default PM level 2 cycles service interval. The default PM
level 2 cycles interval is 2 times that of the PM level 1 cycles interval.