3
Routine Maintenance
12
STERRAD
®
Sterilization System Routine Maintenance Guide
Maintenance Schedule
PM1 PM2
Task
Description
Yes
Yes
Inspect Seismic Restraints
Yes
Yes
Replace Vacuum Pump Oil
Yes
Yes
Vaporizer Plate and Vaporizer Maintenance
Yes
Yes
Drain Water From Air Storage Tank
Yes
Yes
Drain Pressure Regulator Moisture Trap
Yes
Yes
Replace Vacuum Pump Oil Filter (Leybold D16A only)
Yes
Yes
Clean Air Filters/Panel Assemblies
No
Yes
Baratron Zero Shift Test
No
Yes
Replace the Catalytic Converter
No
Yes
Replace HEPA Filter
Yes
Yes
Replace Exhaust Filters
Yes
Yes
Clean Door and Inside Chamber Assembly
No
Yes
Clean Platen Surface
No
Yes
Clean Door Seal and Replace Door Seal O-ring
No
Yes
Inspect throttle valve and O-ring. Clean if needed.
No
Yes
Replace Upper Shelf Stops
No
Yes
Replace Electrode Spacers
Yes Yes Product
Certification
Yes Yes Update
PM
Log