1.5.
Technical Data SMC 4
Standard configuration
Objective: GF - Planachromat 1x
Working distance 100 mm
Eyepiece
GF- Pw 10x /20
Tube factor
2x
Zoom - Objective
Zoom 10:1
Selected factors at magnification changer
0,32x ... 3,2x
Total
magnification
V
t
= V
obj
x V
eyepiece
x V
tube
x V
changer
6,3x ... 63x
Object field (mm)
32 … 3,2
Depth of field (mm)
5,1 … 0,08
Adjustable range of coarse drive 45mm
Max. High of objects
135 mm
Interpupillary adjustment 55...80 mm
Adjustment ametropia
+/- 6 dpt
Dimension in mm
320 x 320 x 415
(W x D x H)
Weight
7 kg
2. Starting Operations
2.1. Assembly
Please open carefully the packaging of the mi-
croscope.
At first the microscope stand (8) and the col-
umn (11) has to be taken out of the packaging
and has to be put on a plan subsoil. The col-
umn has to be set into the holder for the col-
umn (9). After you have done this, you have to
fix the clamp ring for column at the column
(additional part).
After you have taken the middle part of the mi-
croscope (4) from the packaging, this part has
to be assembled directly at the column (11).
Fix it with help of the screw on the backside of
the drive mechanism (10).
The objective (6) is screwed onto the center of
the middle part of the microscope. Please do
not touch the surfaces of the optical parts of the
(GF-) Planachromat. Before the tube 45° (3)
can be assembled to the middle part of the
microscope, all other supplementary modules
and tubes has to be assembled before.
Following the dust cover will be removed from
the eyepiece cone (2) and the eyepieces (1)
has to be assembled into the cone.
The lamp holder will be tighten at the lumi-
nance ring (5). After that the 3W-LED incident
light illuminator (7) can be clamped at the
lamp holder in the right position and can be
moved in the right direction. By rotating the
collector socket, the size of the illuminated field
will be changed. The incident light illuminator
has to be assembled at the power supply at
the stand base (8) or at a 3W-LED transformer.
Now the brightness can be adjusted by a
voltage regulator which is built-in into the base
(8) or directly by the controller at the trans-
former.
9