28
Problem:
Console does not light up when on:
Console lights up when turned on,
but handpiece does not turn:
Motor slowing down or sluggish:
Endo functions displaying incorrectly:
“Check Motor” warning is displayed:
Correction:
• If Sleep Mode LED’s are blinking, press any button on
Control Panel or press foot pedal to exit Sleep Mode.
• Check console to power connection.
• Check fuse. If blown, replace with 1.6A, 250V
slo-blow fuse.
• Check motor plug connection.
• Check foot pedal connection.
• Depress foot pedal.
• Increase RPM.
• Increase Torque setting
• Check that a file is properly seated in the handpiece
and the latch is closed.
• Check for dirty, under-lubricated handpiece.
• Check if handpiece lubricant is draining into motor.
After lubricating and before autoclaving, stand handpiece
on its base to let excess lubricant drain out.
• Verify that ratio setting matches handpiece ratio. Use
Calibration function.
• Turn power switch off, wait 5 seconds, then turn back on
to reset.
• Check that motor is properly connected to the console.
• Check that the proper motor, AE-250-30, is
connected to the console.
TROUBLESHOOTING: