16
1.
Press, hold, and then release
the CAL/SETUP button. Two
quick beeps will sound and the
following menu will display the
Setup options:
2.
Using the FILE Up/Down buttons, move the
selector-arrow to the desired Setup option
and then press the SELECT button.
Prompts will appear for each option (see
descriptions below).
A. Recall Factory Setup:
Restores the
default factory settings and specifications,
overwriting all previous edits and
customized settings. Press the SELECT
button. The following prompt will display:
Move the selector-arrow to the desired
command and then press the SELECT
button:
• Cancel
- Returns to the previous “Setup
Options” menu without changing any
settings. All previous customized settings
will remain intact.
• YES
- Recalls all factory settings. A tone
will sound and the unit will briefly show the
startup screen before displaying the default
Series S1 settings.
IMPORTANT:
When any of the following
seven Setup Options settings are changed,
the new settings must be saved under the
“Save Settings” command before they are
activated. Otherwise, the last settings used
will stay active. See “Save Settings” option,
page 19, for more details.
Note:
The Setup Options are system-wide
settings and are not saved into any specific
Series when programming.
B. Auto Stop Rev. (Reverse):
There are
three different Auto Stop modes to choose
from. After pressing the SELECT button, the
following menu will display:
• Auto Reverse
- When the Auto Reverse
mode is enabled, rotation of the handpiece
will automatically stop and reverse when
the selected torque limit is reached. When
maintaining activation of the foot pedal, the
System will automatically alternate between
forward and reverse rotation in an attempt
to free the instrument.
When the Auto Reverse mode is selected,
the unit automatically returns to the “Setup
Options” menu and displays “(AUTO)” at
the “Auto Stop Rev.” prompt.
NOTE:
“Auto
Stop Rev. (AUTO)” is the factory default.
• Manual Reverse
- When the Manual
Reverse mode is enabled, forward rotation
of the handpiece will automatically stop
when the selected torque limit is reached.
Upon releasing and then reapplying
pressure to the foot pedal, the handpiece
will rotate in the reverse direction. The
handpiece will continue to rotate in reverse
until the foot pedal is released once again.
Recall Factory Defaults?
> Cancel
Yes
Setup Options:
> Recall Factory Setup
Auto Stop Rev. (AUTO)
Reverse Beeper
(ON)
Torque Warning
(ON)
Accessory Warn
(OFF)
Sleep Mode
(15)
Keypress Beep
(ON)
System Volume
(40%)
Save Settings
Cancel
Auto Stop Reverse:
> Auto Reverse
Manual Reverse
Auto Stop Off
Cancel