27
C.
Prevacuum Wrapped:
• Preconditioning Pulses - 3
• Minimum Temperature - 132°C (270°F).
• Full Cycle Time - 4 minutes
• Minimum Dry Time - 40 minutes
Avoid sharply bending the cord when auto-
claving. Alternatively, wipe down the motor
cord with disinfecting solution, and/or sleeve
the cord between each patient.
CONSOLE
- The exterior of the console may be
cleaned by wiping with a soft cloth moistened
with a mild detergent or a 1:10 bleach solution
(1 part household bleach to 10 parts water).
IMPORTANT:
Use of other cleaning or
disinfecting solutions may damage the console
and may void the warranty.
FOOT PEDAL
- The exterior of the foot pedal
may be cleaned by wiping with a soft cloth
moistened with mild detergent or disinfecting
solution.
NOTE:
Call Aseptico at 1-800-426-5913 for
any questions or clarifications on the above
sterilization procedures.