SETTING UP THE UNIT:
1.
Unpack the console and check
that the Voltage Selector is set
to the proper voltage. Use 115v
for 110-120v 60Hz voltages, and
the 230v for 220-250v 50Hz
voltages. To change the Voltage
Selector, see page 15.
2.
Attach the power cord to
the back of the Console and plug
into a grounded electrical outlet.
3.
Connect the AE-4B-30 motor to
the receptacle on the lower front
panel of the console.
4.
Attach the “E” Type handpiece to
the motor. See Figure 2.
5.
Attach the supplied AE-7P On/Off
Foot Switch to the connector on
the back of the unit marked “Foot
Switch”.
2
POWER
CORD
CONSOLE
MOTOR
FOOT SWITCH
“E” TYPE
HANDPIECE
MOTOR
MOTOR HOLDER
Do not attempt to operate the Endo DTC
®
while configured in the 115V mode with a
230V power source. This could damage
the electrical components and will void
the warranty.
CAUTION
Figure 1
Figure 2