Aseptico AEU-20 Operation And Maintenance Manual Download Page 8

6.

STERILIZATION:

WARNING

- Sterilize the motor

between each patient use.

WARNING 

- Use of a sterilization

method or temperatures other than
what are prescribed may damage
the motor or present a risk of cross-
contamination between patients.

CAUTION 

- Do not soak or

submerge the motor in any liquid.

STERILIZATION PROCEDURE:

Pre-clean

1)

Thoroughly brush off any visible

signs of debris from the AE-4B motor
and cord.

2)

Thoroughly clean the device with a

moist cloth or towel to remove any
remaining signs of debris.

Sterilize

3)

Select one of the three following

sterilization methods (A. B. or C.):

Wrapped Sterilization

– Place in an

appropriately sized sterilization pouch
and seal it.

A. Gravity Wrapped:

• Minimum Temperature:

132° C (270° F) 

• Full Cycle Time:

20 minutes 

• Minimum Dry time:

30 minutes 

B. Prevacuum Wrapped: 

• Preconditioning Pulses: 

• Minimum Temperature:

132° C (270° F)

• Full Cycle Time: 

4 minutes 

• Minimum Dry time:

40 minutes 

Flash Sterilization

– For immediate

use only.

C

Gravity Unwrapped:

• Minimum Temperature:

132° C (270° F) 

• Full Cycle Time:

10 minutes 

No dry time is required for flash
sterilization.

MOTOR & CORD ASSEMBLY:

The entire motor and cord assembly is
fully autoclavable. Before autoclaving
the assembly, remove any handpiece
from motor and install autoclaving plugs
into the end of the motor and motor cord
connector as shown below. The plugs
are supplied with the motor and cord. 

Loosely coil the motor cord when
autoclaving.  Avoid sharply bending the
cord when autoclaving.

NOTE:

Call Aseptico, Inc. at 1-425-426-

5913 for any questions or clarifications
on this sterilization procedure.

•  Always install the autoclaving plugs into motor

and motor connector before autoclaving.

•  If the autoclaving plug for the motor becomes

difficult to remove or install, apply silicone grease
or similar lubricant to the O-rings.

Turn black connector ring only

Summary of Contents for AEU-20

Page 1: ...ENANCE INSTRUCTION MANUAL AEU 20 Endodontic System Endo ITR TM P O Box 1548 Woodinville WA 9 8333 216th Street SE Woodinville WA 98072 International 425 487 3157 Toll Free 800 426 5913 www aseptico co...

Page 2: ...ING Serious injury or death may result if ignored CAUTION Damage to property or the environ ment may result if ignored NOTE Important additional information and hints P O Box 1548 Woodinville WA 98072...

Page 3: ...ries Congratulations This system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico...

Page 4: ...ord to the back of the console and plug into a grounded electrical receptacle 3 Connect the M 8 10K motor to the recepta cle on the lower front panel of the console 4 Attach the appropriate E Type han...

Page 5: ...ion 4 Ratio Selector Calibrates the RPM dis play to read accurately for up to four differ ent handpiece ratios Push the ratio up down arrow until the indicator light matches the handpiece you are usin...

Page 6: ...and operates the motor in a continu ous 360 rotation at maximum torque Operating speed and torque level for each ratio position are stored in the computer memory For most endodontic root canal applic...

Page 7: ...the handpiece being used i e 1 8 for a 1 8 handpiece or 1 16 for a 1 16 handpiece 3 Adjust the up down arrows on the RPM display to the desired operating speed on the RPM readout NOTE The RPM display...

Page 8: ...Pulses 3 Minimum Temperature 132 C 270 F Full Cycle Time 4 minutes Minimum Dry time 40 minutes Flash Sterilization For immediate use only C Gravity Unwrapped Minimum Temperature 132 C 270 F Full Cycl...

Page 9: ...ly with any of the above instructions may void your warranty WARNING HANDPIECES Thorough cleaning and lubrication of handpieces after each use and before sterilization is very important to ensure prop...

Page 10: ...lo fuse for 110 volts 0 16A 250V slo blo fuse for 220 volts Check motor plug connection Check foot switch connection Depress foot switch Increase RPM Check that an instrument is properly seated in the...

Page 11: ...9 NOTES...

Page 12: ...discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to...

Reviews: