Aseptico AEU-20 Operation And Maintenance Manual Download Page 2

TABLE OF CONTENTS:

Introduction  . . . . . . . . . . . . . . . . . . . . . . . . .1

Package Contents . . . . . . . . . . . . . . . . . . . .1

Setting Up the Unit  . . . . . . . . . . . . . . . . . . .2

Operation Functions  . . . . . . . . . . . . . . . . . .3

Operation  . . . . . . . . . . . . . . . . . . . . . . . . . .5

Sterilization  . . . . . . . . . . . . . . . . . . . . . . . . .6

Maintenance & Cleaning . . . . . . . . . . . . . . .7

Symbol Definition  . . . . . . . . . . . . . . . . . . . .7

Troubleshooting  . . . . . . . . . . . . . . . . . . . . .8

Warranty  . . . . . . . . . . . . . . . . . . .Back Cover

SPECIFICATIONS:

Console Size: 5.80” x 7.21” x 3.03”

(14.73 x 18.31 x 7.70 cm)

Console Weight: 3.5 lbs (1.59 kg)

Power Source: 110V/120V, 60 Hz, or

220V/250V, 50 Hz

Fuse:  0.315A - 110V

0.16A - 230V
Slo-Blo Type

CLASSIFICATIONS:

Class I

Type B Applied Parts

Ordinary Protection

Not suitable for use in the presence of a flam-
mable anaesthetic mixture mixed with air or
with oxygen or nitrous oxide.

Note:

The power supply cord is the mains

appliance disconnect means.

To prevent injury to people and damage to property,
please heed relevant warnings and remarks. They
are marked as follows:

WARNING:

Serious injury or death may result if
ignored.

CAUTION:

Damage to property or the environ-
ment may result if ignored.

NOTE:

Important additional information and
hints.

P.O. Box 1548 • Woodinville, WA 98072

8333 216th Street SE • Woodinville, WA 98072

(425) 487-3157  (800) 426-5913

www.aseptico.com • [email protected]

WARNING: This device has been tested and found
to comply with the emissions requirements of IEC
60601-1-2:2001-09.  These requirements provide
reasonable protection against harmful electromag-
netic interference in a typical medical installation.
However, high levels of radio-frequency (RF) emis-
sions from electrical devices, such as cellular
phones, may disrupt the performance of this device.
To mitigate disruptive electromagnetic interference,
position this device away from RF transmitters and
other sources of electromagnetic energy.

MEDICAL EQUIPMENT WITH
RESPECT TO ELECTRIC SHOCK,
FIRE, AND MECHANICAL HAZARDS
ONLY IN ACCORDANCE WITH UL
2601-1 CAN/CSA C22.2 No. 601.1

41EJ

Summary of Contents for AEU-20

Page 1: ...ENANCE INSTRUCTION MANUAL AEU 20 Endodontic System Endo ITR TM P O Box 1548 Woodinville WA 9 8333 216th Street SE Woodinville WA 98072 International 425 487 3157 Toll Free 800 426 5913 www aseptico co...

Page 2: ...ING Serious injury or death may result if ignored CAUTION Damage to property or the environ ment may result if ignored NOTE Important additional information and hints P O Box 1548 Woodinville WA 98072...

Page 3: ...ries Congratulations This system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico...

Page 4: ...ord to the back of the console and plug into a grounded electrical receptacle 3 Connect the M 8 10K motor to the recepta cle on the lower front panel of the console 4 Attach the appropriate E Type han...

Page 5: ...ion 4 Ratio Selector Calibrates the RPM dis play to read accurately for up to four differ ent handpiece ratios Push the ratio up down arrow until the indicator light matches the handpiece you are usin...

Page 6: ...and operates the motor in a continu ous 360 rotation at maximum torque Operating speed and torque level for each ratio position are stored in the computer memory For most endodontic root canal applic...

Page 7: ...the handpiece being used i e 1 8 for a 1 8 handpiece or 1 16 for a 1 16 handpiece 3 Adjust the up down arrows on the RPM display to the desired operating speed on the RPM readout NOTE The RPM display...

Page 8: ...Pulses 3 Minimum Temperature 132 C 270 F Full Cycle Time 4 minutes Minimum Dry time 40 minutes Flash Sterilization For immediate use only C Gravity Unwrapped Minimum Temperature 132 C 270 F Full Cycl...

Page 9: ...ly with any of the above instructions may void your warranty WARNING HANDPIECES Thorough cleaning and lubrication of handpieces after each use and before sterilization is very important to ensure prop...

Page 10: ...lo fuse for 110 volts 0 16A 250V slo blo fuse for 220 volts Check motor plug connection Check foot switch connection Depress foot switch Increase RPM Check that an instrument is properly seated in the...

Page 11: ...9 NOTES...

Page 12: ...discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to...

Reviews: